1.7 Functions of the 6-CH, Medium-Speed Four-
Wire RTD Resistance Input Module
This module allows up to six inputs of DC voltage, 4-wire RTD, 4-wire resistance, and
digital input (DI) at a minimum measurement interval of 100 ms.
Input terminal (clamp terminal)
Terminal cover
Measurement Input Types
Measurement Input Type
Notation
No measurement
SKIP
DC voltage
VOLT
Resistance temperature detector
RTD
Resistance
OHM
DI
DI
Measurement Range
DC Voltage
Measurement Range
Notation
Rated Measurement Range
20 mV
20 mV
–20.000 to 20.000 mV
60 mV
60 mV
–60.00 to 60.00 mV
200 mV
200 mV
–200.00 to 200.00 mV
2 V
2 V
–2.0000 to 2.0000 V
6 V
6 V
–6.000 to 6.000 V
20 V
20 V
–20.000 to 20.000 V
100 V
100 V
–100.00 to 100.00 V
60 mV (high resolution)
60 mVH
0.000 to 60.000 mV
6 V (high resolution)
6 VH
0.0000 to 6.0000 V
1 V
1 V
–1.0000 to 1.0000 V
Resistance Temperature Detector (1 mA)
Measurement Range
Notation
Rated Measurement Range
Pt100
Pt100-1
–200.0 to 600.0°C
JPt100
JPt100-1
–200.0 to 550.0°C
Pt100 (high resolution)
Pt100-1H
–140.00 to 150.00°C
JPt100 (high resolution)
JPt100-1H
–140.00 to 150.00°C
Ni100 SAMA
Ni100SAMA
–200.0 to 250.0°C
Ni100 DIN
Ni100DIN
–60.0 to 180.0°C
Ni120
Ni120
–70.0 to 200.0°C
Pt50
Pt50
–200.0 to 550.0°C
Cu10 GE
Cu10GE
–200.0 to 300.0°C
Cu10 L&N
Cu10LN
–200.0 to 300.0°C
Cu10 WEED
Cu10WEED
–200.0 to 300.0°C
Cu10 BAILEY
Cu10BAILEY
–200.0 to 300.0°C