5.8 4-CH, Medium-Speed Strain Input Module

(MX112) Specifications

Style number:

S2

Number of inputs:

4

Input type:

Strain gauge or strain gauge type sensors (static strain)

Input method:

Floating balanced input isolated between channels (NDIS is non-
isolated)

Measurement range and accuracy:

The accuracy applies to standard operating conditions:

Ambient temp: 23±2°C, ambient humidity: 55±10% RH, supply
voltage: 90 to 250 VAC, power frequency: 50/60 Hz ± 1%, warm-
up time: at least 30 minutes, without adverse conditions such as
vibrations.

One-Gauge Method Conversion

Input

Measurement

Range Type

Rated

Measurement

Range

Integration time: 16.67 ms or more

Integration time: 1.67 ms

Measurement

Accuracy

Resolution

Measurement

Accuracy

Resolution

Strain

2000 µStrain

±2000.0 µStrain

±0.5% of range

0.1 µStrain

±2% of range

1 µStrain

*1

20000 µStrain

±20000 µStrain

±0.3% of range

1 µStrain

±1% of range

2 µStrain

*2

200000 µStrain

±200000 µStrain

±0.3% of range

10 µStrain

±1% of range

10 µStrain

*1

Display resolution is 0.1 µStrain

*2

Display resolution is 1 µStrain

AD resolution:

Equivalent to ± 20000 FS display

However, excludes 1.67 ms integration time

AD integration time:

Measurement

Interval

1.67 ms

16.67 ms

20 ms

Auto

*2

36.67 ms

100 ms

200 ms

*3

200 ms

Integration

Time

Filter

Rejected Noise and Notes

Rectangular

100 ms

200 ms

500 ms

1 s

2 s

5, 10, 20, 30, 60 s

Trapezoidal
Rectangular
Cos

600 Hz and its integer multiples

*1

60 Hz and its integer multiples

50 Hz and its integer multiples

Automatically detects the power supply frequency and set 16.67 or 20 ms
50 Hz, 60 Hz and their integer multiples

10 Hz and its integer multiples
Fc = 5-Hz low-pass filter

*1 When the measurement interval is 100 ms, measured values may fluctuate since power supply frequency

noise is not rejected. In such cases, set the measurement interval to 200 ms or more.

*2 For DC power, set to 20 ms.

*3 When synchronizing time by SNTP, the integral time is set to 100 ms. Also in this case, noise of 10 Hz and its integer

multiples is rejected.

Gauge connection method:

1-gauge (2 or 3 wire systems), opposing 2 gauge, adjacent 2- or
4-gauge

With clamp terminals, set on a channel basis with switches

Applicable gauge resistance: 100 to 1000 Ω

120 Ω for -B12; 350 Ω, built-in resistance for -B35

Bridge voltage:

Fixed at 2 VDC. Accuracy ±5% compensated with internal Cal

Applicable gauge factor:

Fixed at 2.0. Gauge factor can be compensated with the scaling
function

Balance adjustment: Automatic, digital calculation methods
Balance adjustment range:

±10000 µstrain (1 gauge method conversion)

Balance adjustment accuracy:

The measurement accuracy or less

Resistance accuracy for bridge:

±0.01% ±5ppm/°C

Input resistance:

1 MΩ or more.