beautypg.com

Measurement input types, Measurement range, App index – Yokogawa Standalone MW100 User Manual

Page 55

background image

1-41

IM MW100-01E

Explanation of Functions

1

2

3

4

5

App

Index

1.6 Functions of the 30-CH, Medium-Speed DCV/

TC/DI Input Module

This module allows up to thirty inputs of DC voltage, thermocouple, and digital input (DI)
at a shortest measurement interval of 500 ms. It takes up three modules worth of space
when attaching to the base plate.

Input terminal

(M3 screw terminal)

Terminal

cover

Input terminal

(clamp terminal)

Terminal

cover

MX110-VTD-L30/H3

MX110-VTD-L30

Measurement Input Types

Measurement Input Type

Notation

No measurement

SKIP

DC voltage

VOLT

Thermocouple

TC

DI

DI

Remote RJC

RRJC

Measurement Range

DC Voltage

Measurement Range

Notation

Rated Measurement Range

20 mV

20 mV

–20.000 to 20.000 mV

60 mV

60 mV

–60.00 to 60.00 mV

200 mV

200 mV

–200.00 to 200.00 mV

2 V

2 V

–2.0000 to 2.0000 V

6 V

6 V

–6.000 to 6.000 V

20 V

20 V

–20.000 to 20.000 V

100 V

100 V

–100.00 to 100.00 V

60 mV (high resolution)

60 mVH

0.000 to 60.000 mV

6 V (high resolution)

6 VH

0.0000 to 6.0000 V

1 V

1 V

–1.0000 to 1.0000 V

Thermocouple

Measurement Range

Notation

Rated Measurement Range

Type-R

R

0.0 to 1760.0°C

Type-S

S

0.0 to 1760.0°C

Type-B

B

0.0 to 1820.0°C

Type-K

K

–200.0 to 1370.0°C

Type-E

E

–200.0 to 800.0°C

Type-J

J

–200.0 to 1100.0°C

Type-T

T

–200.0 to 400.0°C