App index – Yokogawa Standalone MW100 User Manual
Page 41
1-27
IM MW100-01E
Explanation of Functions
1
2
3
4
5
App
Index
Replacing the CF Card While Recording
You can replace the CF card while the recording is in progress. Replace the CF card
quickly while the access indicator (in-progress display) to the CF card is not ON. If
the time to write the data arrives while you are replacing the CF card, the data for that
interval is dropped. You can check the time when data is written using the recording
status log in the log information.
The measured, computed, or thinned data files that are divided due to the replacement of
the CF card can be joined using the MW100 Viewer Software.
For a description of the CF card access indicator, see “Displays” in this section.
For a description of the time when data is written to the CF card, see appendix 8, “Saving
Data to the CF Card.”
Saving Manual Sample Data
Each time you perform a manual sample, the data is written to the manual sample file
saved on the CF card.
For details on manual samples: see “Manual Sample Function” in this section.
Saving Report Data (/M3 Option)
Hourly, daily, weekly, and monthly data is written to the daily, weekly, and monthly files
saved on the CF card.
For details on reports, see section 1.16, “Report Function (/M3 Option).”
Saving Settings
MW100 setting values can be saved. The contents that are saved are as follows:
• Range, alarm, and MATH related settings
• Media related settings
• Communication related settings
• Other settings
However, when settings are loaded onto the MW100, the IP address, subnet mask,
default gateway, host name, and domain name are not loaded.
For details on the saved items, see saved settings in “Recording Structure” in section 5.2.
Format
Initializes the CF Card.
For information on formatting the CF card, see “Formatting the CF Card and Checking the
Free Space” in section 3.3.
1.3 Functions of the Main Module