Yokogawa Standalone MW100 User Manual
Page 202
4-10
IM MW100-01E
The MW100 can be detected using the Search button, but browser connection fails.
Possible Problem
Corrective Action
Ref. section
The IP address is set to the default value.
The default value cannot be used to make the
connection.
Set the correct IP address.
*
There is a problem in the network configuration.
Check that the IP address, subnet mask, and default
gateway settings of the PC and the MW100 settings are
correct.
*
The PC software, main module style number, and
release number rule is not upheld.
Check the PC software release number and main module
style number before upgrading the version.
[PC software release no.] ≥ [main module style no.]
Page i
*
See the MW100 Viewer Software User’s Manual (IM MW180-01E).
MW100 Calibration Software not connected.
Possible Problem
Corrective Action
Ref. section
Attempting to make multiple connections. Another
software program is already connected.
Exit all other software programs.
4.3, *
The main module and input/output module style
number rule is not upheld.
Check the release number of the PC software and the
main module style number, then upgrade the version.
[PC software release no.] ≥ [main module style no.] ≥ [input/
output module style no.]
Page i
*
See the MW100 Viewer Software User’s Manual (IM MW180-01E).
The connected input/output module is not detected.
Possible Problem
Corrective Action
Ref. section
Module connection or module startup error.
Attached the module while the power was ON.
Turn OFF the power. Detach the input/output module once
and attach it again.
2.5
The PC software, main module, and input/output
module style and release number rule is not
upheld.
Check the PC software release number and main module
style number before upgrading the version.
[PC software release no.] ≥ [main module style no.] ≥
[input/output module style no.]
Page i
Carried out an incorrect calibration.
Recalibrate.
4.3, *
*
See the MW100 Viewer Software User’s Manual (IM MW180-01E).
4.3 Troubleshooting