beautypg.com

Operation modes and statuses, Operation modes and statuses -13, App index – Yokogawa Standalone MW100 User Manual

Page 27

background image

1-13

IM MW100-01E

Explanation of Functions

1

2

3

4

5

App

Index

• In-Progress Display
The following displays cycle while the CF card is being accessed or while calibration

is being performed. Do not remove the CF card while it is being accessed.

For the handling of the CF card, see “Handling of the CF Card” in section 2.10.
For CF card replacement, see “Saving Data to the CF Card” in this section.

• Access Forewarning to the CF Card
When saving measured, computed, or thinned data, the dots blink before accessing

of the CF card. This indication starts 10 seconds before the access. If you see this
indication, quickly finish the insertion or removal of the CF card.

In-progress display

Blinking

If you are using the multi interval function, this indication may be shorter than 10

seconds. If the time until the CF card is accessed is less than or equal to 5 s, the time
until access is displayed numerically.

In-progress display

For CF card replacement, see “Saving Data to the CF Card” in this section.

When performing a manual sample, and when saving report data, the dots do not

blink before accessing of the CF card.

• Non-execution Display
If the file division action is not executed, “--nuLL--” is displayed with the characters

flowing from right to left as shown below. Check the execution condition of the file
division.

Operation Modes and Statuses

The MW100 has a Setting mode in which input ranges and other settings can be entered,
and a Measurement mode in which data acquisition is performed. The mode switches
depending on the measurement item of the status information.

Mode

Status Info

Description

Measurement

Setting mode

Stop

For entering range, system, communication, and

display settings

Measurement mode

Start

For data monitoring, computation, and recording

* The instrument must be in measurement mode in order for the status of computation and

recording to be Start.

1.3 Functions of the Main Module