Measurement input types, Measurement range – Yokogawa Standalone MW100 User Manual
Page 52
1-38
IM MW100-01E
1.5 Functions of the 10-CH, Medium-Speed
Universal Input Module
This module allows up to ten inputs of DC voltage, thermocouple, 3-wire RTD, and digital
input (DI) at a minimum measurement interval of 100 ms.
Input terminal (clamp terminal)
Terminal cover
Measurement Input Types
Measurement Input Type
Notation
No measurement
SKIP
DC voltage
VOLT
Thermocouple
TC
Resistance temperature detector
RTD
DI
DI
Remote RJC
RRJC
Measurement Range
DC Voltage
Measurement Range
Notation
Rated Measurement Range
20 mV
20 mV
–20.000 to 20.000 mV
60 mV
60 mV
–60.00 to 60.00 mV
200 mV
200 mV
–200.00 to 200.00 mV
2 V
2 V
–2.0000 to 2.0000 V
6 V
6 V
–6.000 to 6.000 V
20 V
20 V
–20.000 to 20.000 V
100 V
100 V
–100.00 to 100.00 V
60 mV (high resolution)
60 mVH
0.000 to 60.000 mV
6 V (high resolution)
6 VH
0.0000 to 6.0000 V
1 V
1 V
–1.0000 to 1.0000 V
Thermocouple
Measurement Range
Notation
Rated Measurement Range
Type-R
R
0.0 to 1760.0°C
Type-S
S
0.0 to 1760.0°C
Type-B
B
0.0 to 1820.0°C
Type-K
K
–200.0 to 1370.0°C
Type-E
E
–200.0 to 800.0°C
Type-J
J
–200.0 to 1100.0°C
Type-T
T
–200.0 to 400.0°C
Type-N
N
0.0 to 1300.0°C
Type-W
W
0.0 to 2315.0°C