beautypg.com

Measurement input types, Measurement range – Yokogawa Standalone MW100 User Manual

Page 52

background image

1-38

IM MW100-01E

1.5 Functions of the 10-CH, Medium-Speed

Universal Input Module

This module allows up to ten inputs of DC voltage, thermocouple, 3-wire RTD, and digital
input (DI) at a minimum measurement interval of 100 ms.

Input terminal (clamp terminal)

Terminal cover

Measurement Input Types

Measurement Input Type

Notation

No measurement

SKIP

DC voltage

VOLT

Thermocouple

TC

Resistance temperature detector

RTD

DI

DI

Remote RJC

RRJC

Measurement Range

DC Voltage

Measurement Range

Notation

Rated Measurement Range

20 mV

20 mV

–20.000 to 20.000 mV

60 mV

60 mV

–60.00 to 60.00 mV

200 mV

200 mV

–200.00 to 200.00 mV

2 V

2 V

–2.0000 to 2.0000 V

6 V

6 V

–6.000 to 6.000 V

20 V

20 V

–20.000 to 20.000 V

100 V

100 V

–100.00 to 100.00 V

60 mV (high resolution)

60 mVH

0.000 to 60.000 mV

6 V (high resolution)

6 VH

0.0000 to 6.0000 V

1 V

1 V

–1.0000 to 1.0000 V

Thermocouple

Measurement Range

Notation

Rated Measurement Range

Type-R

R

0.0 to 1760.0°C

Type-S

S

0.0 to 1760.0°C

Type-B

B

0.0 to 1820.0°C

Type-K

K

–200.0 to 1370.0°C

Type-E

E

–200.0 to 800.0°C

Type-J

J

–200.0 to 1100.0°C

Type-T

T

–200.0 to 400.0°C

Type-N

N

0.0 to 1300.0°C

Type-W

W

0.0 to 2315.0°C