beautypg.com

Switches and keys, Switches and keys -11, App index switches and keys – Yokogawa Standalone MW100 User Manual

Page 25: User function keys, Dip switch 1, Dip switch 2, Key lock function, Explanation of functions, Firmware update, 3 functions of the main module

background image

1-11

IM MW100-01E

Explanation of Functions

1

2

3

4

5

App

Index

Switches and Keys

The MW100 has the following switches and keys. Some are included with options.
• Start and Stop keys
• User function key 1
• User function key 2
• Dip switch 1
• Dip switch 2
• Terminator switch (/C3 option)
• Power switch

User Function Keys

Actions set up using the Event/Action function can be executed by pressing the user
function keys on the front panel of the MW100.
The keys are assigned as follows by default.

Key

Display

Action

User function key 1

USER1

Write to setting values file

User function key 2

USER2

Load setting values file

Dip Switch 1

Used to initialize the MW100 settings and for other functions.
• Normal operation

• Initialization of IP addresses and other settings

1

2

3

4

5

6

7

8

ON

1

2

3

4

5

6

7

8

ON

• Fixed IP address (192.168.0.10) • 10-Mbps half-duplex Ethernet communication

1

2

3

4

5

6

7

8

ON

1 2 3 4 5 6 7 8

ON

• Firmware update

1 2 3 4 5 6 7 8

ON

1 2 3 4 5 6 7 8

ON

Main unit

Web

Dip Switch 2

Turn all switches ON for normal operation. If the switches are set differently, the
instrument may not function correctly.

1

2

ON

Key Lock Function

You can apply a lock to the functions of the Start, Stop, and user function keys. The lock
prevents inadvertent execution of functions.

For setting the key lock, see “Status Information and Processing/Operation” in section 3.3.

1.3 Functions of the Main Module