beautypg.com

4 gauge method – Yokogawa Standalone MW100 User Manual

Page 104

background image

2-16

IM MW100-01E

• 4 Gauge Method

ON

OFF

No.1

No.2

No.3

No.4

No.5

A(+V)

B( L)

C(-V)

D( H)

-B12, -B35

No.1

OFF

No.2

OFF

No.3

OFF

No.4

OFF

No.5

ON

Rg1

Rg2

Rg4

Rg1

Rg2

Rg3

E

e

Rg1, Rg3

Rg2, Rg4

Rg1

Rg3

Rg1

Rg3

Rg4

Rg2

Rg1, Rg2

Rg3, Rg4

Rg3

Rg4

R: fixed resistance

r: resistance value of lead wire

Rg: resistance value of strain gauge

e: output voltage from bridge

E: voltage applied to bridge

Setting switch

-NDI

Setting switch

Rg1

Rg2

Rg4

Rg3

ON

1 2 3 4 5

1

2

3

4

5

6

7

8

OFF
SW

Bridge head

(701955 or 701956)

SW1

OFF

SW2

OFF

SW3

OFF

SW4

ON

SW5

OFF

2.4 Connecting Signal Wires