beautypg.com

App index – Yokogawa Standalone MW100 User Manual

Page 260

background image

5-47

IM MW100-01E

Specification

1

2

3

4

5

App

Index

Pt500

*2

Pt1000

*2

–200.0 to 600.0°C

–200.0 to 600.0°C

0.000 to 20.00 Ω

0.00 to 200.00 Ω

0.0 to 2000.0 Ω

Vth = 2.4 V

±(0.05% of rdg + 3digits)

±(0.05% of rdg + 7digits)

±(0.05% of rdg + 3digits)

±(0.05% of rdg + 3digits)

Threshold level accuracy: ±0.1 V

±(0.1% of rdg + 1.5digits)

±(0.1% of rdg + 25digits)

±(0.1% of rdg + 15digits)

±(0.1% of rdg + 10digits)

Input

Measurement Accuracy

Integration Time: 16.67 ms

or More

Rated

Measurement

Range

Measurement Accuracy

Integration Time: 1.67 ms

Measurement

Range

Type

Highest

Resolution

(1 Digit)

0.1°C

0.001 Ω

0.01 Ω

0.1 Ω

Resistance

DI

RTD

*3

(Measurement

current: 0.25

mA)

*1 Measured using a measurement current of approximately 10 μA at 200 mV range. Threshold level is approximately 0.1 V.

*2 The Pt500 resistance table is Pt100 × 5, and the resistance table for Pt1000 is Pt100 × 10

*3 4-wire RTD, 4-wire resistance.

ON for 100 Ω or less, OFF for 10 kΩ or more

*1

20 Ω

(Measurement

current 1 mA)

2 kΩ

(Measurement

current 0.25 mA)

Level

Contact input

200 Ω

(Measurement

current 1 mA)

Measurement interval, integration time, and filter:

Measurement

Interval

1.67 ms

16.67 ms

20 ms

Auto

*2

36.67 ms

100 ms

*3

200 ms

*4

200 ms

Integration

Time

Filter

Rejected Noise and Notes

Rectangular

100 ms

200 ms

500 ms

1 s

2 s

5 s

10, 20, 30, 60 s

Trapezoidal
Rectangular
Cos

600 Hz and its integer multiples

*1

60 Hz and its integer multiples

50 Hz and its integer multiples

Automatically detects the power supply frequency and set 16.67 or 20 ms

50 Hz, 60 Hz and their integer multiples

10 Hz and its integer multiples
Fc = 5-Hz low-pass filter

*1 When the measurement interval is 100 ms or 200 ms, measured values may fluctuate (especially for temperature,

20 Ω, and other measurements) since power supply frequency noise is not rejected. In such cases, set the

measurement interval to 500 ms or more.

*2 For DC power, set to 20 ms.

*3 When synchronizing time by SNTP, the integral time is set to 36.67 ms. Also in this case, noise of 50 Hz, 60 Hz, and

their integer multiples is rejected.

*4 When synchronizing time by SNTP, the integral time is set to 100 ms. Also in this case, noise of 10 Hz and its integer

multiples is rejected.

Maximum input voltage:

DC voltage at 1-V range or less, RTD, resistance, and DI (contact):
±10 VDC (continuous)

Other measurement ranges: ±120 VDC (continuous)

Normal-mode voltage:

DC voltage, DI (LEVEL): 1.2 times the range rating or less (50/60
Hz, peak value including the signal component)

Resistance 2 kΩ, RTD 100 Ω, 500 Ω: 1000 Ω types: 50 mVpeak

Resistance 200 Ω, RTD 10 Ω, 25 Ω: 50 Ω types: 10 mVpeak

Resistance 20 Ω: 4 mVpeak

Normal-mode rejection ratio:

For integration time of 16.67 ms or more: 40 dB or more (50/60
Hz ±0.1%)

For integration time of 1.67 ms: 50/60 Hz is not rejected

Common-mode voltage: 600 VACrms (50/60 Hz), reinforced (double) insulation
Common mode rejection ratio (50/60 Hz ±0.1%, 500 Ω unbalanced between minus

measurement terminal and ground. The RTD and resistance
ranges indicate the voltage conversion value when current flows)

When the integration time is 16.67 ms or more: 120 dB or more

When the integration time is 1.67 ms: 80 dB or more

Common-mode voltage between channels:

DC voltage, DI: 120 VACrms (50/60 Hz)

RTD, resistance: 50 VACrms (50/60 Hz)

5.7 6-CH, Medium-Speed Four-Wire RTD Resistance Input Module (MX110-V4R-M06) Specifications