beautypg.com

Effects of operating conditions, Effects of operating conditions -48 – Yokogawa Standalone MW100 User Manual

Page 261

background image

5-48

IM MW100-01E

Noise rejection:

Rejection by the integrating A/D converter and the use of low pass
filters

Input resistance:

For DC voltage 1 V range or less: 10 MΩ or less

For DC voltage 2 V range or higher: Approx. 1 MΩ

Insulation resistance: Between input terminal and earth terminal: 20 MΩ or more (500

VDC)

Input bias current:

10 nA or less

Withstand voltage:

1000 VACrms (50/60 Hz) for one minute between input terminals
(DC voltage and DI)

620 VACrms (50/60 Hz) for one minute between input terminals
(RTD and resistance)

3700 VACrms (50/60 Hz) for one minute between the input
terminal and earth terminal

Input signal source resistance:

DC voltage: 2 kΩ or less

Resistance and RTD ranges: 10 Ω or less per line (same for all
ranges)

Shunt capacity:

0.01 µF or less (when using RTD and resistance ranges)

Power consumption: Approx. 1.2 W
External dimensions: Approx. 57 (W) × 131 (H) × 151 (D) mm (including the terminal

cover)

Weight:

Approx. 0.5 kg.

Terminal type:

Clamp, terminal board is detachable

Applicable wire size: 0.14 to 1.5 mm

2

(AWG26 to 16)

Effects of Operating Conditions

The specifications below apply when the integration time is 16.67 ms or more.
Warm-up time:

At least 30 minutes after power-up

Effects of ambient temperature:

The effect received by changes in 10 degree increments of the
ambient temperature is ±(0.05% of rdg. + 0.05% of range) or less.
However, for Cu10 Ω: ±(0.2% of range + 1 digit)

Effects of power fluctuation:

Meets the accuracy specifications for AC power supply in the
range 90 to 132 V and 180 to 250 V

Effects of magnetic field: The fluctuation in external magnetic fields for AC (50/60 Hz)

400 A/m is ±(0.1% of rdg. + 10 digits) or less

Effects from the signal source resistance: Effect from fluctuation in DC voltage signal

source resistance of 1 kΩ

1 V range or less

±10 µV or less

2 V range or higher

±0.15% of rdg. or less

RTD:

Fluctuation from 10 Ω change per line

1000 Ω, 100 Ω types: ± 0.1°C or less

Other than 1000 Ω, 100 Ω types: ± 1.0°C or less

Resistance: Fluctuation from 10 Ω change per line: ±1 digit or less

Effect of position:

Horizontal with the feet at the bottom is the rule.

Effects of vibration:

The fluctuation that results by applying a sinusoidal vibration
along all three axes at a frequency between 10 to 60 Hz and an
acceleration of 0.2 m/s

2

is ±(0.1% of rdg. + 1 digit) or less

5.7 6-CH, Medium-Speed Four-Wire RTD Resistance Input Module (MX110-V4R-M06) Specifications