Saving measured data and computed data, Saving thinned data, Saving data at multi intervals – Yokogawa Standalone MW100 User Manual
Page 40: File division, File message
1-26
IM MW100-01E
Saving Measured Data and Computed Data
Files can be created for every measurement group. An individual file is created for
computed data. For each measurement group, you can select whether or not to perform
the save operation.
The table below shows the approximate interval over which data can be saved to the CF
card when one measurement interval is used.
Save Channel
Measurement Interval
Capacity of the CF Card
128 MB
512 MB
10 ch
10 ms
8.8 hours
35.3 hours
100 ms
3.7 days
14.8 days
1 s
37 days
148 days
Saving Thinned Data
Data is saved at a specified thinning interval. Thinning time:
4, 5, 10, 20, or 30 seconds, or 1, 2, 3, 4, 5, 10, 20, or 30 minutes, or 1 hour.
Select a thinning time from the above choices.
Saving Data at Multi Intervals
The recording operation can be specified for each interval group. Multiple sets of data
can be saved simultaneously such as recording an interval around a sudden event while
continuously recording long-term changes in the data.
Time (s)
Start
Trigger 1
condition met
Interval group 1: Trigger (pretrigger 50%)
Interval group 2: Direct
Interval group 3: Trigger
Thinning:
Trigger 1
condition met
Trigger 3
condition met
File Division
You can use the Event/Action function, communication commands, or operations in the
browser’s status information screen to divide the file containing measured, computed, or
thinned data at an arbitrary time. After dividing a file, the next file division is enabled 10
minutes later.
Action
Notation
Operation and Notes
Save recorded data
Memory Save
Divides and saves the measured or computed data file.
in divisions
This action is valid when the recording start action is set
to Direct and the recording stop action is set to Fullstop
or Rotate.
Save thinning
Memory Save (T) Divides and saves the thinned data file.
recording in divisions
This action is valid when the recording stop action is set
to Fullstop or Rotate.
File Message
You can enter a file message of up to 120 characters common to the measured and
computed data file, and a message for thinned data file. You can view the file message
on the MW100 Viewer Software.
1.3 Functions of the Main Module