Saving measured data and computed data, Saving thinned data, Saving data at multi intervals – Yokogawa Standalone MW100 User Manual

Page 40: File division, File message

background image

1-26

IM MW100-01E

Saving Measured Data and Computed Data

Files can be created for every measurement group. An individual file is created for
computed data. For each measurement group, you can select whether or not to perform
the save operation.
The table below shows the approximate interval over which data can be saved to the CF
card when one measurement interval is used.

Save Channel

Measurement Interval

Capacity of the CF Card

128 MB

512 MB

10 ch

10 ms

8.8 hours

35.3 hours

100 ms

3.7 days

14.8 days

1 s

37 days

148 days

Saving Thinned Data

Data is saved at a specified thinning interval. Thinning time:
4, 5, 10, 20, or 30 seconds, or 1, 2, 3, 4, 5, 10, 20, or 30 minutes, or 1 hour.
Select a thinning time from the above choices.

Saving Data at Multi Intervals

The recording operation can be specified for each interval group. Multiple sets of data
can be saved simultaneously such as recording an interval around a sudden event while
continuously recording long-term changes in the data.

Time (s)

Start

Trigger 1

condition met

Interval group 1: Trigger (pretrigger 50%)
Interval group 2: Direct
Interval group 3: Trigger
Thinning:

Trigger 1

condition met

Trigger 3

condition met

File Division

You can use the Event/Action function, communication commands, or operations in the
browser’s status information screen to divide the file containing measured, computed, or
thinned data at an arbitrary time. After dividing a file, the next file division is enabled 10
minutes later.

Action

Notation

Operation and Notes

Save recorded data

Memory Save

Divides and saves the measured or computed data file.

in divisions

This action is valid when the recording start action is set

to Direct and the recording stop action is set to Fullstop

or Rotate.

Save thinning

Memory Save (T) Divides and saves the thinned data file.

recording in divisions

This action is valid when the recording stop action is set

to Fullstop or Rotate.

File Message

You can enter a file message of up to 120 characters common to the measured and
computed data file, and a message for thinned data file. You can view the file message
on the MW100 Viewer Software.

1.3 Functions of the Main Module