beautypg.com

Warning – Yokogawa Standalone MW100 User Manual

Page 110

background image

2-22

IM MW100-01E

Wiring the Power Supply Terminal (When the Suffix Code of the Power Supply/Cord Is
-2
c* or -3W)

* c is D, F, R, Q, or H.

When using an AC adapter for the power supply

WARNING

• To prevent the possibility of electric shock when wiring, confirm that the power

supply source is turned OFF.

• Use the power cord supplied by Yokogawa with the instrument.
• Check that the supply voltage matches the rated voltage of the AC adapter

before connecting the power cord.

• During periods of extended non-use, remove the AC adapter’s power cord from

the AC outlet.

• Do not use an AC adapter (model:772075) other than the one supplied by

YOKOGAWA.

• Nothing should be placed on top of the AC adapter power cord. The power cord

should also be kept away from any heat sources.

• When unplugging the power cord from the outlet, never pull by the cord itself.

Always hold and pull by the plug. If the power cord is damaged, contact your
dealer for replacement.

Use a power supply that meets the conditions indicated below.

Item

Specifications

Rated supply voltage

100 to 240 VACrms

Supply voltage range used

AC power supply, 90 to 250 VACrms

Rated supply voltage frequency 50/60 Hz
Allowable line frequency range 50/60 Hz ±2%
Maximum power consumption

Up to approximately 70 VA maximum when six modules are used

1.

Check that the power supply and the power switch of the main module is OFF.

2.

Loosen the screw fixing the power supply terminal cover of the main module in

place and open the power supply terminal cover.

3.

Connect the plus (red) and minus (blue) wires from the AC adapter and the

protective ground wire to the power supply terminal as in the figure below.

4.

Close the power supply terminal cover and secure it with the screw.

DATA ACQUISITION UNIT

12 - 28V DC

Protective

grounding line

Power supply

terminal cover

12 - 28V DC

Minus wiring

Minus wiring

Plus wiring

Plus wiring

Protective

grounding line

2.5 Connecting the Power Supply and Turning the Power Switch ON and OFF