Execution Errors
The code is divided into two parts which are displayed alternately on the 7-segment LED;
in the first part, the letter E appears in the left digit with the hundreds digit of the error
code to the right, and the second part consists of the last two digits of the error code.
Display
Possible Problem
Corrective Action
E201
Cannot execute due to different operation mode. Confirm the operation mode.
E202
Cannot execute when in setting mode.
Change the mode before execution.
E203
Cannot execute when in measurement mode.
Change the mode before execution.
E204
Cannot change or execute during memory sampling. Stop the save operation before executing.
E205
Cannot execute during MATH operation.
Stop the MATH operation before executing.
E206
Cannot change or execute during MATH operation. Stop the MATH operation before executing.
E207
Cannot change or execute while saving/loading
Execute after the settings are saved or loaded.
settings.
E209
Cannot execute while memory sample is stopped. Change the mode before execution.
E211
No relays for communication input found.
Check installation of relays and the relay output types.
E212
Initial balance failed.
Check the settings and wiring.
E213
No channels for initial balance found.
Check the target channels.
E214
No channels for transmission output found.
Specify channels for transmission output.
E215
No channels for arbitrary output found.
Specify channels for arbitrary output.
E221
No measurement channels found.
Check the measurement module, measurement group number,
measurement interval, and other settings.
E222
Invalid measurement interval.
Set the measurement interval so that: Meas. gr 1 ≤ meas. gr 2
≤ meas. gr 3
E223
Too many measurement channels.
The number of measurable channels during 10 ms
measurement is 10, and for 50 ms measurement, 30.
E224
No MATH channels found.
Check the MATH channel settings.
E225
Invalid MATH interval.
Set the MATH interval to measurement groups of 100 ms or
more. When measuring with measurement modules, set the
measurement group numbers on which to perform
measurement.
E226
Cannot start/stop MATH operation.
Cannot execute because MATH start is set for the level
detection action.
E227
Cannot start/stop recording.
Cannot execute because recording start is set for the level
detection action.
4.1 Error Display on the 7-Segment LED and Corrective Actions