Communication errors, System errors, Communication errors -7 system errors -7 – Yokogawa Standalone MW100 User Manual
Page 199: App index communication errors
4-7
IM MW100-01E
Troubleshooting and Maintenance
1
2
3
4
5
App
Index
Communication Errors
The code is divided into two parts which are displayed alternately on the 7-segment LED;
in the first part, the letter E appears in the left digit with the hundreds digit of the error
code to the right, and the second part consists of the last two digits of the error code.
Display
Possible Problem
Corrective Action
E501
Login first.
First, finish logging in.
E502
Login failed, try again.
Enter the correct user name and password.
E503
Connection count exceeded the upper limit.
Close unneeded connections and reconnect.
E504
Connection has been lost.
Try to make a new connection.
E505
Connection has time out.
Try to make a new connection.
E520
FTP function not available.
Enable the function.
E521
FTP control connection failed.
Check the FTP server address and the main unit address
setting. Also check the Ethernet cable connection.
E530
SMTP function not available.
Enable the function.
E531
SMTP connection failed.
Check the SMTP server address and the main unit address
setting. Also check the Ethernet cable connection.
E532
POP3 connection failed.
Check the POP3 server address and the main unit address
setting. Also check the Ethernet cable connection.
E550
SNTP function not available.
Enable the function.
E551
SNTP command/response failed.
Check the SNTP server address and the main unit address
setting. Also check the Ethernet cable connection.
System Errors
The code is divided into two parts which are displayed alternately on the 7-segment LED;
in the first part, the letter E appears in the left digit with the hundreds digit of the error
code to the right, and the second part consists of the last two digits of the error code.
Display
Possible Problem
Corrective Action
E999
System error.
Servicing required.
4.1 Error Display on the 7-Segment LED and Corrective Actions