Measurement Range
Notation
Rated Measurement Range
J263B
J263B
0.0 to 300.0K
Cu10 at 20°C
Cu10a392
–200.0 to 300.0°C
alpha=0.00392
Cu10 at 20°C
Cu10a393
–200.0 to 300.0°C
alpha=0.00393
Cu25 at 0°C
Cu25
–200.0 to 300.0°C
alpha=0.00425
Cu53 at 0°C
Cu53
–50.0 to 150.0°C
alpha=0.00426035
Cu100 at 0°C
Cu100
–50.0 to 150.0°C
alpha=0.00425
Pt25 (JPt100 × 1/4)
Pt25
–200.0 to 550.0°C
Cu10 GE (high resolution)
Cu10GEH
–200.0 to 300.0°C
Cu10 L&N (high resolution)
Cu10LNH
–200.0 to 300.0°C
Cu10 WEED (high resolution)
Cu10WEEDH
–200.0 to 300.0°C
Cu10 BAILEY (high resolution)
Cu10BAILEYH
–200.0 to 300.0°C
Pt100 GOST
Pt100G
–200.0 to 600.0°C
Cu100 GOST
Cu100G
–200.0 to 200.0°C
Cu50 GOST
Cu50G
–200.0 to 200.0°C
Cu10 GOST
Cu10G
–200.0 to 200.0°C
Resistance Temperature Detector (0.25 mA)
Measurement Range
Notation
Rated Measurement Range
Pt500
Pt500
–200.0 to 600.0°C
Pt1000
Pt1000
–200.0 to 600.0°C
Resistance
Measurement Range
Notation
Rated Measurement Range
20 Ω (measured current 1 mA)
20 ohm
0.000 to 20.000 Ω
200 Ω (measured current 1 mA)
200 ohm
0.00 to 200.00 Ω
2 kΩ (measured current 0.25 mA)
2000 ohm
0.0 to 2000.0 Ω
DI
Measurement Range
Notation
Rated Measurement Range
LEVEL
LEVEL
Vth=2.4V
Contact input
CONTACT
1 kΩ or less, ON, 100 kΩ or less, OFF
(shunt capacitance: 0.01 µF or less)
Measurement Interval, Integration Time, and Filter
You can select from the following measurement intervals for this module.
100 ms, 200 ms, 500 ms, 1 s, 2 s, 5 s, 10 s, 20 s, 30 s, 60 s
The integral time and types of filters applied vary depending on the measurement
interval.
For information on filters, see section 2.9, “Measures Against Noise on the MW100 Data
Acquisition Unit.”
Measurement Synchronization
Since measurement is sequential, measurement on each channel is not synchronized.
1.7 Functions of the 6-CH, Medium-Speed Four-Wire RTD Resistance Input Module