Effects of operating conditions, General specifications, External dimensions – Yokogawa Standalone MW100 User Manual
Page 258: App index
5-45
IM MW100-01E
Specification
1
2
3
4
5
App
Index
Up/Down setting is possible
2 kΩ or less normal, 200 kΩ or more disconnected
(shunt capacitance: 0.01 μF or less), detection current approx. 10
μA, detection time approx. 1.6 ms
Power consumption: Approx. 1.2 W
External dimensions: Approx. 174 (W) × 131 (H) × 150 (D) mm (including the terminal cover)
Weight:
Approx. 0.8 kg.
Terminal type:
Clamp, or M3 screw terminal (if /H3 option is added)
Applicable wire size: 0.14 to 1.5 mm
2
(AWG26 to 16) (for clamp terminals)
Effects of Operating Conditions
The specifications below apply when the integral time is 16.67 ms or more.
Warm-up time:
30 minutes after power -up.
Effects of ambient temperature:
The effect received by changes of 10°C increments of the ambient
temperature is
±(0.05% of rdg. + 0.05% of range) or less
Effects of power fluctuation:
Meets the accuracy specifications for AC power supply in the
range 90 to 132 V and 180 to 250 V.
Effect of magnetic field: The fluctuation in external magnetic fields for AC (50/60 Hz) 400
A/m is ±(0.1% of rdg. + 10 digits) or less
Effects from the signal source resistance:
Effect from fluctuation in DC voltage and thermocouple signal
source resistance of 1 kΩ
DC voltage:
1 V range or lower ±10 μV or less
2 V range or higher ±0.15% of rdg. or less
Thermocouple: ±10 μV or less
Effect of position:
Horizontal with the feet at the bottom is the rule.
Effects of vibration:
The fluctuation that results by applying a sinusoidal vibration
along all three axes at a frequency between 10 to 60 Hz and an
acceleration of 0.2 m/s
2
is ±(0.1% of rdg. + 1 digit) or less.
General Specifications
Operating temperature range:
–20 to 60°C
Operating humidity range:
20 to 80% RH for –20 to 40°C
10 to 50% RH for 40 to 50°C
5 to 30% RH for 50 to 60°C
External Dimensions
Units: mm
131
174
150
If not specified, the tolerance is ±3%. However, in cases of less than 10 mm, the
tolerance is ±0.3 mm.
5.6 30-CH, Medium-Speed DCV/TC/DI Input Module (MX110-VTD-L30) Specifications