beautypg.com

Effects of operating conditions, General specifications, External dimensions – Yokogawa Standalone MW100 User Manual

Page 258: App index

background image

5-45

IM MW100-01E

Specification

1

2

3

4

5

App

Index

Up/Down setting is possible

2 kΩ or less normal, 200 kΩ or more disconnected

(shunt capacitance: 0.01 μF or less), detection current approx. 10
μA, detection time approx. 1.6 ms

Power consumption: Approx. 1.2 W
External dimensions: Approx. 174 (W) × 131 (H) × 150 (D) mm (including the terminal cover)
Weight:

Approx. 0.8 kg.

Terminal type:

Clamp, or M3 screw terminal (if /H3 option is added)

Applicable wire size: 0.14 to 1.5 mm

2

(AWG26 to 16) (for clamp terminals)

Effects of Operating Conditions

The specifications below apply when the integral time is 16.67 ms or more.
Warm-up time:

30 minutes after power -up.

Effects of ambient temperature:

The effect received by changes of 10°C increments of the ambient
temperature is

±(0.05% of rdg. + 0.05% of range) or less

Effects of power fluctuation:

Meets the accuracy specifications for AC power supply in the
range 90 to 132 V and 180 to 250 V.

Effect of magnetic field: The fluctuation in external magnetic fields for AC (50/60 Hz) 400

A/m is ±(0.1% of rdg. + 10 digits) or less

Effects from the signal source resistance:

Effect from fluctuation in DC voltage and thermocouple signal
source resistance of 1 kΩ

DC voltage:

1 V range or lower ±10 μV or less

2 V range or higher ±0.15% of rdg. or less

Thermocouple: ±10 μV or less

Effect of position:

Horizontal with the feet at the bottom is the rule.

Effects of vibration:

The fluctuation that results by applying a sinusoidal vibration
along all three axes at a frequency between 10 to 60 Hz and an
acceleration of 0.2 m/s

2

is ±(0.1% of rdg. + 1 digit) or less.

General Specifications

Operating temperature range:

–20 to 60°C

Operating humidity range:

20 to 80% RH for –20 to 40°C

10 to 50% RH for 40 to 50°C

5 to 30% RH for 50 to 60°C

External Dimensions

Units: mm

131

174

150

If not specified, the tolerance is ±3%. However, in cases of less than 10 mm, the
tolerance is ±0.3 mm.