beautypg.com

Yokogawa Standalone MW100 User Manual

Page 200

background image

4-8

IM MW100-01E

4.2 Error Display in the Monitor Screen and

Corrective Actions

Error Message

Corrective Action

Could not connect to the instrument.

Check Ethernet cable connections and the IP addresses of devices.

Check cables and other connections.

Communication error occurred.

Check Ethernet cable connections and the IP addresses of devices.

Check cables and other connections.

The actually installed modules differ from the modules

Reconstruct the module configuration.

recognized by the system.
The size of the data files exceeds the allowable upper limit. Set the number of save channels, recording interval, and recording

data length so that the data files of measurement groups 1, 2, and 3,

and the computed and thinned data files are all within 10 MB.

Value smaller than measuring interval or MATH interval

Set a value higher than the measurement and MATH interval.

cannot be set for the thinning recording interval.
Combination of thinning recording interval, measurement Set the thinning recording interval so that it is a common multiple of

interval, MATH interval incorrect.

the measurement and MATH intervals.

The recording data length cannot be set equal to or less Set a recording data length larger than the recording interval.

than the recording interval.
Insufficient space on the CF card.

Delete unneeded files on the CF card to free up some space.

Replace the CF card.

The CF card is not inserted.

Insert the CF card.

The CF card is damaged or not formatted.

Reinsert the CF card or format it.

No data found in file.

Check the recording settings.

Remove MATH start/stop action from the Event/Action

Remove MATH start/stop action from the Event/Action settings.

settings.
Remove recording start/stop action from the Event/Action Remove recording start/stop action from the Event/Action settings.

settings.