Reference Junction Compensation of Thermocouple Input

Normally, the input terminal of the input module is nearly at room temperature,
therefore the actual thermocouple output differs from the value in the table for the
thermoelectromotive force with the 0°C standard. Modules able to measure temperature
with thermocouples can be compensated by measuring the temperature of the input
terminal and adding the corresponding thermoelectromotive force to the actual
thermocouple output. Therefore with the measurement terminal shorted (equivalent to
an edge detection of 0°C), the measured value indicates the temperature of the input
terminal.
When calibrating modules capable of temperature measurements using thermocouples,
it is necessary to apply input from a DC standard voltage current generator with this
compensation voltage subtracted (the electromotive force of the 0°C standard that is
equivalent to the temperature of the input terminal). As in the figure, when performing
reference junction compensation at 0°C using a 0°C standard temperature device, you
can perform the calibration by inputting the 0°C standard electromotive force from DC
standard voltage/current generator.

Wiring Diagram

+

-

100Vpk MAX TO

250V MAX CH TO CH

600V MAX TO

Thermocouple wires or

TC extension wires

Thermocouple

wires

Copper

wires

Copper wires

DC voltage standard

Input terminal

+

Cold junction

0°C standard temperature

device

4-CH, High-Speed Universal

Input Module*

* Connect the 10-CH, Medium Speed

Universal Input Module and the

30-CH, Medium Speed DCV/TC/DI

Input Module in the same manner

as well.

Note

• Calibration of temperature measurements of the MW100 Data Acquisition Unit using

thermocouples differs from calibration of DC voltage and RTD ranges in that the input
cannot be adjusted. If the temperature measurement calibration using thermocouples does
not meet the accuracy specifications, check thoroughly for input error and other problems,
then contact your Yokogawa dealer or representative.

• If errors exist in the thermocouple wires and TC extension wires, correct calibration is not

possible. Be sure to use a calibrated thermocouple.

4.4 Calibration