Yokogawa Standalone MW100 User Manual
Page 259
5-46
IM MW100-01E
5.7 6-CH, Medium-Speed Four-Wire RTD
Resistance Input Module (MX110-V4R-M06)
Specifications
Style number:
S2
Type of measurement: DC voltage, 4-wire RTD, 4-wire resistance, DI (contact, LEVEL)
Number of inputs:
6
Input method:
Floating unbalanced input, insulation between channels
A/D resolution:
16 bits (±20000/±6000/0 to 60000)
Measurement range and accuracy:
The accuracy applies to standard operating conditions:
Ambient temp: 23 ±2°C, ambient humidity: 55 ±10% RH, supply
voltage: 90 to 250 VAC, power frequency: 50/60 Hz ±1%, warm-
up time: at least 30 minutes, without adverse conditions such as
vibrations.
20 mV
60 mV
200 mV
2 V
6 V
20 V
100 V
60 mV (high res.)
6 V (high res.)
1 V
Pt100
*1
JPt100
*1
Pt100 (high res.)
JPt100 (high res.)
Ni100 SAMA
*2
Ni100 DIN
*2
Ni120
*3
Pt50
Cu10 GE
*4
Cu10 L&N
*4
Cu10 WEED
*4
Cu10 BAILEY
*4
J263B
Cu10 at 20°C
alpha=0.00392
Cu10 at 20°C
alpha=0.00393
Cu25 at 0°C
alpha=0.00425
Cu53 at 0°C
alpha=
0.00426035
Cu100 at 0°C
alpha=0.00425
Pt25
(JPt100/4)
Cu10 GE
(high resolution)
Cu10 L&N
(high resolution)
Cu10 WEED
(high resolution)
Cu10 BAILEY
(high resolution)
Pt100 GOST
Cu100 GOST
Cu50 GOST
Cu10 GOST
DC voltage
–20.000 to 20.000 mV
–60.00 to 60.00 mV
–200.00 to 200.00 mV
–2.0000 to 2.0000 V
–6.000 to 6.000 V
–20.000 to 20.000 V
–100.00 to 100.00 V
0.000 to 60.000 mV
0.0000 to 6.0000 V
–1.0000 to 1.0000 V
–200.0 to 600.0°C
–200.0 to 550.0°C
–140.00 to 150.00°C
–140.00 to 150.00°C
–200.0 to 250.0°C
–60.0 to 180.0°C
–70.0 to 200.0°C
–200.0 to 550.0°C
–200.0 to 300.0°C
–200.0 to 300.0°C
–200.0 to 300.0°C
–200.0 to 300.0°C
0.0 to 300.0K
–200.0 to 300.0°C
–200.0 to 300.0°C
–200.0 to 300.0°C
–50.0 to 150.0°C
–50.0 to 150.0°C
–200.0 to 550.0°C
–200.0 to 300.0°C
–200.0 to 300.0°C
–200.0 to 300.0°C
–200.0 to 300.0°C
–200.0 to 600.0°C
–200.0 to 200.0°C
–200.0 to 200.0°C
–200.0 to 200.0°C
±(0.05% of rdg + 0.3°C)
±(0.05% of rdg + 0.3°C)
±(0.05% of rdg + 0.3°C)
±(0.1% of rdg + 2°C)
±(0.05% of rdg + 0.3K)
±(0.1% of rdg + 2°C)
±(0.1% of rdg + 0.5°C)
±(0.05% of rdg + 0.3°C)
±(0.1% of rdg + 0.5°C)
±(0.1% of rdg + 2°C)
±(0.05% of rdg + 0.3°C)
±(0.05% of rdg + 0.3°C)
±(0.05% of rdg + 0.3°C)
±(0.1% of rdg + 2°C)
±(0.1% of rdg + 25 digits)
±(0.1% of rdg + 10 digits)
±(0.1% of rdg + 100 digits)
±(0.1% of rdg + 100 digits)
±(0.1% of rdg + 10 digits)
±(0.1% of rdg + 1.5°C)
±(0.1% of rdg + 1.5°C)
±(0.1% of rdg + 1.5°C)
±(0.2% of rdg + 5°C)
±(0.1% of rdg + 1.5K)
±(0.2% of rdg + 5°C)
±(0.2% of rdg + 2°C)
±(0.1% of rdg + 1.5°C)
±(0.2% of rdg + 2°C)
±(0.2% of rdg + 5°C)
±(0.1% of rdg + 1.5°C)
±(0.1% of rdg + 1.5°C)
±(0.1% of rdg + 1.5°C)
±(0.2% of rdg + 5°C)
Input
Measurement Accuracy
Integration Time: 16.67 ms
or More
Rated
Measurement
Range
Measurement Accuracy
Integration Time: 1.67 ms
Measurement
Range
Type
*1 Pt50: JIS C 1604-1989, JIS C 1606-1989/Pt100: JIS C 1604-1997, IEC 751, DIN IEC 751/JPt100: JIS C 1604-1989, JIS C 1606-1989
*2 SAMA/DIN
*3 McGRAW EDISON COMPANY
*4 Guaranteed accuracy range Cu10 GE: –84.4 to 170.0°C/Cu10 L&N: –75.0 to 150.0°C/Cu10 WEED: –20.0 to 250.0°C/
Cu10 BAILEY: –20.0 to 250.0°C
*5. 4-wire RTD, 4-wire resistance
Highest
Resolution
(1 digit)
1 µV
10 µV
10 µV
100 µV
1 mV
1 mV
10 mV
1 µV
100 µV
100 µV
0.1°C
0.01°C
0.1°C
0.1°C
0.1K
0.1°C
0.1°C
0.1°C
0.1°C
0.1°C
0.1°C
0.1°C
0.1°C
0.1°C
RTD
*5
(Measurement
current: 1 mA)
±(0.05% of rdg + 5 digits)
±(0.05% of rdg + 2 digits)
±(0.05% of rdg + 2 digits)
±(0.05% of rdg + 5 digits)
±(0.05% of rdg + 2 digits)
±(0.05% of rdg + 2 digits)
±(0.05% of rdg + 2 digits)
±(0.05% of rdg + 20 digits)
±(0.05% of rdg + 20 digits)
±(0.05% of rdg + 2 digits)