3.4 Setting Acquisition Conditions for Measured/

Computed Data

This is an explanation of measurement settings, MATH settings, recording settings, and
thinning settings.

Setting Mode

Measurement Operation Settings

1.

From the top screen, click

System Setting > Measurement Setting.

Measurement Group Settings

2.

Select the measurement interval from the

Interval list. Set a measurement

interval for each group number.

Measurement Module Settings

Sets the integral time and measurement group assignments for each module.

3.

Select the group you wish to assign in the

Interval Group list.

4.

Select an integral time of AUTO, 50 Hz, or 60 Hz in the

A/D Integration Time list.

This setting is enabled from the measurement interval.

5.

Click the

Apply button. The setting changes take effect.

For a description of the measurement interval and integration time that can be specified, see

section 2.9, “Measures Against Noise on the MW100 Data Acquisition Unit.”

Note

• The measurement interval that is assigned to the measurement group number is,

(interval is short) meas. group1 ≤ meas. group2 ≤ meas. group 3 (interval is long)

• The equivalent of three modules worth of settings are entered for the 30-CH Medium Speed

DCV/TC/DI Input Module.
• Select the same measurement group for the three measurement groups.
• Select the same integral time for the three A/D integral times.