3-20

IM MW100-01E

Computation Operation Settings

From the top screen, click

System Setting > MATH Setting.

MATH Interval Settings

1.

In the

Interval Group list, select the interval group number. The measurement

interval is set to the interval assigned to the selected group number.

Note

If multiple interval groups are specifying the same measurement interval, specify the largest
number for the interval group number that you select.

MATH Operation Settings

2.

Select +Over or –Over in the

Error Data list.

TLOG and CLOG Computation Operation Settings

3.

Select units of integration in the

Sum Scale list.

4.

Select Over (computation stop) or Rotate in the

PSUM Overflowed Data list.

Measurement Input Settings

5.

Select Error (MATH error) or Skip in the

Abnormal Input Data list.

6.

Select Over (handle as an overflow value), Skip, or Limit in the

Overflowed Data

list.

TLOG and CLOG Input Settings

7.

Select Error (MATH error) or Skip in the

Abnormal Input Data list.

8.

Select Error (MATH error), Skip, or Limit in the

Overflowed Data list.

9.

Click the

Apply button. This enables the settings changes.

For processing upon MATH errors, see “Processing Computed Results with Abnormal Input

Values or Overflow Values” in section 1.15.

3.4 Setting Acquisition Conditions for Measured/Computed Data