beautypg.com

Starting and stopping computation, Starting and stopping recording, App index starting and stopping computation – Yokogawa Standalone MW100 User Manual

Page 169

background image

3-47

IM MW100-01E

Setting and Data acquisition

1

2

3

4

5

App

Index

Starting and Stopping Computation

This operation is available when the MATH function (/M1 option) is included, or when the
10-CH Pulse Input Module is installed. You can use the function when MATH channels
are set.
Neither the computation start nor the computation stop operation can be done while
recording MW100. It is possible to operate it from the Event/Action function and a
browser. However, even if computation start/computation stop is set by the Event/Action
function, it is disregarded while MW100 is not recording.

Using Main Module Keys
• Computation Start

While recording is stopped, and during measurement, briefly press the

START key.

Computation starts.

• Computation Stop

During computation, and while recording is stopped, briefly press the

STOP key.

Computation stops.

Using a Browser (in the Setting Screen)

The main module can be switched to Measurement mode. This is done while recording is
stopped.

• Starting Computation

In the

Status table in the top screen, select Start from the Operation list of the MATH

item. Computation starts.

• Stopping Computation

In the

Status table in the top screen, select Stop from the Operation list of the MATH

item. Computation stops.

You can also start or stop the computation from the Monitor screen.

For the procedure to start or stop the computation in the Monitor screen, see section 3.16,

“Measured Data Monitor Display/Settings.”

Starting and Stopping Recording

Using Main Module Keys
• Starting Recording

During measurement, hold down the

START key for two or more seconds. Recording

starts.

• Stopping Recording

During recording, hold down the

STOP key for two or more seconds. Recording stops.

Using a Browser (in the Setting Screen)

The main module can be switched to Measurement mode.

• Starting Recording

In the

Status table in the top screen, select Start from the Operation list of the

Recording item. Recording starts.

• Stopping Recording

In the

Status table in the top screen, select Stop from the Operation list of the

Recording item. Recording stops.

You can also start or stop the recording from the Monitor screen.

For the procedure to start or stop the recording in the Monitor screen, see section 3.16,

“Measured Data Monitor Display/Settings.”