3-47
IM MW100-01E
Setting and Data acquisition
1
2
3
4
5
App
Index
Starting and Stopping Computation
This operation is available when the MATH function (/M1 option) is included, or when the
10-CH Pulse Input Module is installed. You can use the function when MATH channels
are set.
Neither the computation start nor the computation stop operation can be done while
recording MW100. It is possible to operate it from the Event/Action function and a
browser. However, even if computation start/computation stop is set by the Event/Action
function, it is disregarded while MW100 is not recording.
Using Main Module Keys
• Computation Start
While recording is stopped, and during measurement, briefly press the
START key.
Computation starts.
• Computation Stop
During computation, and while recording is stopped, briefly press the
STOP key.
Computation stops.
Using a Browser (in the Setting Screen)
The main module can be switched to Measurement mode. This is done while recording is
stopped.
• Starting Computation
In the
Status table in the top screen, select Start from the Operation list of the MATH
item. Computation starts.
• Stopping Computation
In the
Status table in the top screen, select Stop from the Operation list of the MATH
item. Computation stops.
You can also start or stop the computation from the Monitor screen.
For the procedure to start or stop the computation in the Monitor screen, see section 3.16,
“Measured Data Monitor Display/Settings.”
Starting and Stopping Recording
Using Main Module Keys
• Starting Recording
During measurement, hold down the
START key for two or more seconds. Recording
starts.
• Stopping Recording
During recording, hold down the
STOP key for two or more seconds. Recording stops.
Using a Browser (in the Setting Screen)
The main module can be switched to Measurement mode.
• Starting Recording
In the
Status table in the top screen, select Start from the Operation list of the
Recording item. Recording starts.
• Stopping Recording
In the
Status table in the top screen, select Stop from the Operation list of the
Recording item. Recording stops.
You can also start or stop the recording from the Monitor screen.
For the procedure to start or stop the recording in the Monitor screen, see section 3.16,
“Measured Data Monitor Display/Settings.”
3.13 Starting and Stopping Measurement, Computation, and Recording