2.7 Connecting the RS-422A/485 Interface (/C3

Option)

Terminal Wiring and Signal Names

SERIAL COMM

FG

SDB SDA RDB RDA

SG

FG (Frame Ground)

Case grounding for the main unit.

SG (Signal Ground)

The signal ground.

SDB (Send Data B)

Send data B(+).

SDA (Send Data A)

Send data A(–).

RDB (Received Data B)

Received data B(+).

RDA (Received Data A)

Received data A(–).

Connection Procedure

Cables Used

There are two types of cable, a 4-wire and 2-wire cable. Choose a cable depending on
the following conditions.

Cable Used

Twisted pair shielded cable
3 × 24 AWG or more (4-wire), 2 × 24 AWG or more (2-wire)

Characteristic impedance 100 Ω
Capacitance

50 pF/m

Cable length

Max 1.2 km*

* The transmission distance of the RS-422A/485 interface is not the direct distance, but rather

the total cable length (shielded, twisted pair).

WARNING

To prevent electric shock, confirm that the power supply is turned OFF before
making connections.

Connections with Upstream Devices

The following figure shows a connection with an upstream device. If the upstream
devices use an RS-232 port, connect through a converter.

Main unit

RS-422A/485

terminals

Host computer

or connected

upstream instrument

Connected

upstream

instrument

Converter

RS-422A/485

MW100

Main unit

RS-422A/485

terminals

Host computer

RS-232

RS-422A/485

MW100

Connected upstream instrument