5.11 8-CH, Medium-Speed Analog Output Module

(MX120-VAO-M08) Specifications

Style number:

S2

Number of outputs:

8

Update interval:

100 ms minimum (not synchronized to the measurement interval)

Output types:

DC voltage, DC current (external 24-V power supply required
when using current output)

Rated output range: Voltage: –10 to 10 V

Current: 0 to 20 mA sourcing (for 1 to 5 V output, 4 to 20 mA is
output)

Maximum allowable output range:

Voltage: –11 to 11 V Current: 0 to 22 mA

Load impedance:

Voltage: 5 kΩ or more

Current: 600 Ω or less.

Accuracy:

±0.2% of F.S. at the rated output range or less (F.S. = 10 V or
20 mA). However, for current output, accuracy is met at 1 mA or
more.

The accuracy applies to standard operating conditions:

Ambient temp: 23 ±2°C, ambient humidity: 55 ±10% RH, supply
voltage: 90 to 250 VAC, power frequency: 50/60 Hz ±1%, warm-
up time: at least 30 minutes, without adverse conditions such as
vibrations.

Output resolution:

12 bits of F.S. or more

Resolution –10.000 V to 10.000 V (1 mV resolution)

0.000 mA to 20.000 mA (1µA resolution)

Effects of ambient temperature:

±(50 ppm of Setting + 50 ppm of F.S.) per degree 1°C or less
(F.S. = 10 V or 20 mA)

External power supply: 24 V ±10% (required when using current output)

Connect a device with capacity of 250 mA or more.

Insulation resistance: Between output terminals and earth terminal: 20 MΩ or more

(500 VDC)

Across output terminals: non-isolated (- terminal common
potential)

Withstand voltage:

2300 VAC (50/60 Hz) for one minute between the output terminal
and earth.

Across output terminals: non-isolated (-terminal common
potential).

Power consumption: Approx. 2.5 W (not including power consumption of external

power supply).

Terminal type:

Clamp, attached and removed in units of 4 channels

Applicable wire size: 0.08 to 2.5 mm

2

(AWG28 to 12)

External dimensions: Approx. 57 (W) × 131 (H) × 151 (D) mm (including the terminal

cover)

Weight:

Approx. 0.5 kg.

General Specifications

Operating temperature range:

–20 to 50°C

Operating humidity range:

20 to 80% RH for –20 to 40°C

10 to 50% RH for 40 to 50°C