General specifications, General specifications -56 – Yokogawa Standalone MW100 User Manual
Page 269
5-56
IM MW100-01E
5.11 8-CH, Medium-Speed Analog Output Module
(MX120-VAO-M08) Specifications
Style number:
S2
Number of outputs:
8
Update interval:
100 ms minimum (not synchronized to the measurement interval)
Output types:
DC voltage, DC current (external 24-V power supply required
when using current output)
Rated output range: Voltage: –10 to 10 V
Current: 0 to 20 mA sourcing (for 1 to 5 V output, 4 to 20 mA is
output)
Maximum allowable output range:
Voltage: –11 to 11 V Current: 0 to 22 mA
Load impedance:
Voltage: 5 kΩ or more
Current: 600 Ω or less.
Accuracy:
±0.2% of F.S. at the rated output range or less (F.S. = 10 V or
20 mA). However, for current output, accuracy is met at 1 mA or
more.
The accuracy applies to standard operating conditions:
Ambient temp: 23 ±2°C, ambient humidity: 55 ±10% RH, supply
voltage: 90 to 250 VAC, power frequency: 50/60 Hz ±1%, warm-
up time: at least 30 minutes, without adverse conditions such as
vibrations.
Output resolution:
12 bits of F.S. or more
Resolution –10.000 V to 10.000 V (1 mV resolution)
0.000 mA to 20.000 mA (1µA resolution)
Effects of ambient temperature:
±(50 ppm of Setting + 50 ppm of F.S.) per degree 1°C or less
(F.S. = 10 V or 20 mA)
External power supply: 24 V ±10% (required when using current output)
Connect a device with capacity of 250 mA or more.
Insulation resistance: Between output terminals and earth terminal: 20 MΩ or more
(500 VDC)
Across output terminals: non-isolated (- terminal common
potential)
Withstand voltage:
2300 VAC (50/60 Hz) for one minute between the output terminal
and earth.
Across output terminals: non-isolated (-terminal common
potential).
Power consumption: Approx. 2.5 W (not including power consumption of external
power supply).
Terminal type:
Clamp, attached and removed in units of 4 channels
Applicable wire size: 0.08 to 2.5 mm
2
(AWG28 to 12)
External dimensions: Approx. 57 (W) × 131 (H) × 151 (D) mm (including the terminal
cover)
Weight:
Approx. 0.5 kg.
General Specifications
Operating temperature range:
–20 to 50°C
Operating humidity range:
20 to 80% RH for –20 to 40°C
10 to 50% RH for 40 to 50°C