Yokogawa Standalone MW100 User Manual
Page 143
3-21
IM MW100-01E
Setting and Data acquisition
1
2
3
4
5
App
Index
Measurement/Computation Recording Operation Settings
1.
From the top screen, click
System Setting > Recording Setting.
Set the recording operation for each measurement group.
2.
Select the recording start action in the
Mode list under Interval Group.
If you select Direct, steps 3 and 4 are required. If you select Trigger, steps 3 to 6 are
required.
3.
Select the recording stop action in the
Action list under Interval Group.
4.
Select the recording interval in the
Recording Interval list under Interval Group.
The recording interval is set to an integer multiple of the measurement interval. In addition,
the selectable recording interval varies depending on the measurement interval assigned to
the selected interval group.
5.
Select a recording data length from the
Data Length list under Interval Group.
6.
Select a pretrigger length (%) from the
Pre-trigger list under Interval Group.
For the selectable recording intervals, see “Recording of Measured/Computed Values”
under “Recorder Structure” in section 5.2.
Recording Operation Settings
7.
Select a recording data length from the
DIRECT Data Length list. Measurement
groups for which the recording start timing is Direct are all set to this.
8.
Select the remaining space time in the
Limit for Media Alarm list. When the
specified remaining amount is reached, you can have output such as relay output
occur.
For a description of the recording start action, see “Saving Data to the CF Card” in section 1.3.
File Message Settings
9.
Enter the message to be saved to the measured/computed data file in the File
Message box.
10.
Click the
Apply button. The setting changes take effect.
3.4 Setting Acquisition Conditions for Measured/Computed Data