Chapter 5 specification, 1 common specifications, Normal operating conditions – Yokogawa Standalone MW100 User Manual
Page 214: Transport and storage conditions, Mechanical specifications (excluding ac adapter), Standards compliance, App index
5-1
IM MW100-01E
Specification
1
2
3
4
5
App
Index
5.1 Common Specifications
Normal Operating Conditions
Rated power supply voltage: AC power supply (with or without AC adapter): 100 to 240 VAC
DC power supply: 12 to 28 VDC
Supply voltage range used: AC power supply (with or without AC adapter): 100 to 250
VAC
DC power supply: 10 to 32 VDC
Power supply frequency:
50 Hz ±2%, 60 Hz ±2%
Power consumption:
AC power, or DC power with AC adapter: Approx. 70 VA
max.
DC power supply: Approx. 35 VA max.
* When any six I/O modules are installed
Vibration:
10 to 60 Hz 0.2m/s
2
or less
Shock:
Not allowed
Magnetic field:
400 A/m or less (50/60 Hz)
Position:
Position horizontally with feet down
Structure:
Not explosion-proof
Usage location:
Indoors
Operating altitude:
2,000 m or less
Overvoltage category:
II (per IEC61010-1 CSA22.2 No. 61010-1)
Measurement category:
II (per IEC61010-1 CSA22.2 No. 61010-1)
Pollution degree:
2 (per IEC61010-1 CSA22.2 No. 61010-1)
Transport and Storage Conditions
Environmental conditions during transport and storage:
Ambient storage temperature : –25 to 70°C
Ambient storage humidity:
5% to 95% RH (no condensation)
Vibration:
10 to 60 Hz 4.9m/s
2
or less
Shock:
392 m/s
2
or less (packaged condition)
Mechanical Specifications (Excluding AC Adapter)
External dimensions: Approx. 455 Ч 131 Ч 159 mm (when six slots are attached)
Weight:
Approx. 4.3 kg (maximum total weight)
Installation method: Desktop or floor, panel mount, or attached to a DIN rail
Material:
Steel plate, aluminum die-cast, molded plastic resin
Standards Compliance
CSA:
CSA22.2 No 61010-1, overvoltage category II, measurement
category II, pollution degree 2
UL:
Conforms to UL61010B-1 (CSA NRTL/C)
CE:
EMC directive;
EN61326 Class A
EN61000-3-2
EN61000-3-3
Low voltage directive; EN61010-1; overvoltage category II,
measurement category II,
pollution degree 2
C-Tick:
Conforms with AS/NZS CISPR11 Class A Group1
Overvoltage category II: An index for defining transient overvoltages (includes the
rated impulse withstand voltage, and applies to electrical
equipment supplied with power from fixed installations such as
a distribution board).
Pollution degree 2:
The degree of adherence by a solid, liquid, or vapor that
reduces the withstand voltage or surface resistance factor
(applies only to normal indoor atmospheres (with non-
conductive pollution).
Measurement category II: For measurements performed on circuits directly connected
to the electrical distribution system such as provided by a wall
outlet.
Chapter 5
Specification