Configuring the primary data acquisition device – Measurement Computing DBK Part 2 User Manual
Page 88

DBK48, pg. 8
967792
8B Isolated Signal Conditioning Module
Configuring the Primary Data Acquisition Device
DaqBook/100 Series & /200 Series and DaqBoard [ISA type] Configuration
Use of a DBK48 with a DaqBook/100 Series, /200 Series devices, or with an ISA-type DaqBoard requires
the configuration of jumpers JP1 and JP4
.
These jumpers are located on the DaqBook/100 Series, /200
Series devices, and DaqBoard [ISA type] board.
1. If not using auxiliary power, set the JP1 jumper for Analog Option Card Use,
also referred to as the expanded analog mode.
The JP1 default position (above) is necessary to power the interface circuitry of the
DBK48 via the internal ±15 VDC power supply. If using auxiliary power (e.g.,
DBK32A or DBK33) you must remove both JP1 jumpers.
For additional information refer to Power Requirements in the DBK Basics section and
to the DBK32A and DBK33 sections, as applicable.
2. For DaqBook/100, DaqBook /112, and DaqBook /120 only, place the JP4 jumper in the single-ended
mode.
DaqBook/2000 Series, DaqBoard/2000 Series, DaqLab, and DaqScan
No jumper configurations are required on these Daq devices in regard to connecting a DBK48.
LogBooks
No jumper configurations are required on LogBook devices in regard to connecting a DBK48.
Required Jumper Settings in DaqBook/100 Series & /200 Series and ISA-Type DaqBoards
Note:
These jumpers do not
apply to /2000 Series
Devices.