background image

Channel Filter Jumper Settings for DBK50 and DBK51

The low-pass filters have a default corner frequency of 3.5 Hz when the jumpers are in the LPF-selected
positions. This frequency may be readily changed by installing a different value of SIP resistor network in
the 6-pin SIP socket of each filter section. Each channel has its own SIP located next to the channel filter
bypass jumper and labeled RN(1-8)01A. The table lists values of common networks and their corner
frequencies.

Corner Frequency

R-SIP

Bournes Part #

7500 Hz

47 Ω

4606M-102-470

3500 Hz

100 Ω

4606M-102-101

1750 Hz

200 Ω

4606M-102-201

750 Hz

470 Ω

4606M-102-471

350 Hz

1 k Ω

4606M-102-102

175 Hz

2 k Ω

4606M-102-208

75 Hz

4.7 k Ω

4606M-102-208

35 Hz

10 k Ω

4606M-102-103

17.5 Hz

20 k Ω

4606M-102-203

7.5 Hz

47 k Ω

4606M-102-473

3.5 Hz

100k Ω

4606M-102-104

DaqBook/100 Series & /200 Series and DaqBoard [ISA type] Configuration

Several setup steps in DaqBook/100 Series & /200 Series devices and DaqBoards [ISA type] are required
to use a DBK50 or DBK51 module in a system.

1.

If not using auxiliary power, set the JP1 jumper [in the DaqBook/100 Series & /200 Series

devices or ISA-Type DaqBoard] to the Analog Option Card Use, also referred to as the
expanded analog mode.

SSH

DAC 0 - EXT

DAC 0 - INT

DAC 1 - INT

DAC 1 - EXT

JP2

JP1

-15 V

-OCTOUT
-OCLKIN

+15 V

Analog Option

Card Use

UNI-

BI-

SE

16CH

DIFF

8CH

JP3

JP4

Bipolar

Single-Ended

DaqBook/DaqBoard Jumpers for DBK50/51

DBK Option Cards and Modules

989594

DBK50 and DBK51, pg. 3