DBK43A & DBK43B, pg. 32

899892

DBK Option Cards and Modules

Note:

Maximum gain calibration is x1000 for +5V range.

The majority of the gain should be assigned to the Input Amplifier, with the Scaling Amplifier used for
fine-tuning. If the filter is enabled, a gain of x2 is automatically introduced.

The input amplifier has a gain range of ×100 to ×1250; the filter gain ×1 or ×2; and the scaling amplifier
has a range of ×1 to ×10. For the strain gage example, if we round off our gain to ×420, any of these
possible settings will work.

Option A

Option B

Option C

Option D

Input Gain

×420 ×100 ×240

×300

Filter Gain (enabled)

No

Yes (×2)

Yes (×2) *

See Note

No

Scaling Gain

×1 ×2.1 ×1.75

×1.4

Total Gain

×420 ×420 ×420

×420

For Option C, the LPF gain is typically x2.
For gains of x1 (if the filter is enabled), the following apply:

DBK16

- For a gain of x1 (if the filter is enabled),10K

Ω resistors R44 and R46 must have been

previously removed (for the low and high channels, respectively).

DBK43A and DBK43B

- For a gain of x1 (if the channel filters are enabled), removal of the

following 10 K

Ω resistors applies: Ch0 – R144, Ch1 – R244, Ch3 – R444, Ch4 – R544,

Ch5 – R644, Ch6 – R744, Ch7 – R844.