beautypg.com

Measurement Computing DBK Part 2 User Manual

Page 159

background image

DBK Option Cards and Modules

989494

DBK81, DBK82, and DBK83 pg. 9

DBK81, DBK82, DBK83 - Specifications

Name/Function:

DBK81 – 7 Channel High-Accuracy Thermocouple Card

DBK82 – 14 Channel High-Accuracy Thermocouple Card

DBK83 – 14 Channel High-Accuracy Thermocouple Card with external TC/mV

screw-terminal connection pod

System Connector: All DBK options have a DB37 male, which mates with P1 on the DaqBoard, DaqBook,

LogBook, or other DBK options
TC/mV Connector

DBK81: Board-mounted screw terminals

DBK82: Board-mounted screw terminals

DBK83: External pod-mounted screw terminals

Functions: TC types J, K, S, T, E, B, R, N; x100 (voltage)
Inputs

DBK81: 7 differential TC/mV inputs

DBK82: 14 differential TC/mV inputs

DBK83: 14 differential TC/mV inputs

Input Voltage Range: ±100 mV with a DaqBoard/2000 or LogBook

±50 mV with a DaqBook or DaqBoard

Input Impedance: 40M Ohm (differential); 20M Ohm (single-ended)
Input Bandwidth: 4 Hz
Input Bias Current: 10 nA typ
CMRR: 100dB typ
Maximum Working Voltage (signal + common mode): ±10 V
Over-Voltage Protection: ±40 V
Power Requirements

DBK81: 35 mA max from ±15V; 2 mA max from +5 V

DBK82 and DBK83: 60 mA max from ±15V; 2 mA max from +5 V

Operating Temperature: 0°C to 50°C

Voltage Accuracy: ±(0.2% of reading +50 µV)
TC Accuracy: See table and accuracy conditions. Valid for one year, 18 to 28°C
Minimum Resolution: 0.1°C for all TC types

TC Accuracy

at Measurement Temperature in °C (±°C)

Type

Min

Max

-100

0

100

300

500

700

900

1100

1400

J

-200 760 0.8

0.7

0.7

0.8

0.9

0.9

— — —

K

-200 1200 0.9

0.8

0.8

0.9

1.1

1.1

1.2

1.3

T

-200 400 0.9

0.8

0.8

0.8

— — — — —

E

-270 650 0.8

0.7

0.7

0.7

0.8

— — — —

S

-50 1768 — 3.1

2.4

2.0

2.0

1.9

2.0

2.1

2.1

R

-50 1768 — 3.1

2.1

2.0

1.9

1.9

1.7

1.9

2.0

B

50

1780

— — — 4.9

3.2

2.8

2.4

2.3

2.0

N28

-270 400 1.2

0.9

0.9

0.9

— — — — —

N14

0 1300 — 0.9

0.9

0.9

1.1

1.1

1.2

1.3

Accuracy conditions:

Data is based on the use of a calibrated DaqBoard/2000

The table reflects total system absolute accuracy, including accuracy of the CJC and DaqBoard/2000

Excludes possible error from thermocouples

Excludes

noise

VCM = 0

This manual is related to the following products: