DBK Option Cards and Module

899892

DBK43A & DBK43B, pg. 29

Selecting “Mode = Excitation” for DBK Parameter 1

Note

: The Param2 column will show two switch positions for DBK43B applications.

6. Click the Download button to send the current configuration to the LogBook.

7. Select Indicators \ Enable Input Reading Column from the menu bar to display the

excitation values for each channel.

Selecting “Enable Input Reading Column”

(from the Indicators Pull-Down Menu)

8. Set the excitation voltage for each transducer by adjusting the trimpot labeled EXC for the

associated channel while reading their values in LogView.

9. Select Indicators \ Disable Input Reading Column from the menu bar.

Selecting “Disable Input Reading Column”(from the Indicators Pull-Down Menu)

10. DBK43A users: Return the physical calibration switches (of the applicable DBK43As) to the

NORM position.
DBK43B users: Return CAL1 to RUN (Down); and CAL2 to CAL (Up).