beautypg.com

Measurement Computing DBK Part 2 User Manual

Page 178

background image

pg. 4, DBK90

907492

DBK Option Cards and Modules

Open Thermocouple Detection

The DBK90 is equipped with open thermocouple detection for each channel. This means that a broken
thermocouple wire [or otherwise unconnected input] that is measured will result in an off-scale reading.
This is accomplished by applying a small bias current to each of the channel inputs. Whenever a valid
input is absent, the bias current saturates the input amplifier, resulting in the off-scale reading.

Module Configuration

Up to sixteen DBK90 modules can be attached to a single DaqBook/2000 Series device, WBK40, or
WBK41. Since multiple modules are connected via a parallel interface, each must have a unique channel
address.

CAUTION

Adjustment of the channel address must only be performed when the system
power is OFF. Failure to do so may result in equipment damage.

To assign a channel address to the DBK90 module, first locate the DIP switch on the unit’s underside (the
side opposite of the mini-TC connectors). Four micro-switches [on the DIP switch] are used to set the
module’s channel address in binary. After ensuring that the system power is OFF, adjust the micro-
switches to set the desired address. The following figure shows DIP switch settings for the 16 possible
addresses.

DBK90 Channel Address Settings

Each module in the system must have a unique address.

This manual is related to the following products: