beautypg.com

Hardware adjustment, Trimpots, Cal/norm, cal1/run, and cal2/run switches – Measurement Computing DBK Part 2 User Manual

Page 33

background image

DBK Option Cards and Module

899892

DBK43A & DBK43B, pg. 13

2. For DaqBook/100, DaqBook/112 and DaqBook/120, place the JP4 jumper in single-ended mode

Note: To use a DBK43A [or DBK43B] with a Daq PC-Card, you must an appropriate power module must

be used.

DaqBook/2000 Series and DaqBoard/2000 Series Configuration

No Jumper configurations are required for these /2000 series devices.

Hardware Adjustment

Bridge circuit transducers are used for many different applications, and the DBK43A and DBK43B
modules are flexible enough to support most of them. Each channel circuit has an excitation regulator, a
high gain (100-1250) input amplifier with offset adjustment, a low-pass filter, a scaling (1-10) amplifier,
and a calibration multiplexer.

Trimpots

The front of the strain gage modules have a slot to allow access to 4 potentiometers to trim (adjust) the
accuracy for each channel circuit. The trimpots are labeled to represent the following adjustments:

EXC

for adjusting the excitation voltage to the transducer

GAIN

for setting the gain of the input amplifier

OFFSET for adjusting the circuit offset for quiescent loads or bridge imbalance

SCALE for setting the gain of the scaling amplifier

Channel Number

Trimpot

CH0

CH1

CH2

CH3

CH4

CH5

CH6

CH7

EXC

TP101 TP201 TP301 TP401 TP501 TP601 TP701 TP801

GAIN

TP104 TP204 TP304 TP404 TP504 TP604 TP704 TP804

OFFSET

TP103 TP203 TP303 TP403 TP503 TP603 TP703 TP803

SCALE

TP105 TP205 TP305 TP405 TP505 TP605 TP705 TP805

The following figure shows trimpot locations for both the DBK43A and DBK43B.

Trim Pot Locations

CAL/NORM, CAL1/RUN, and CAL2/RUN Switches

DBK43A has a single CAL/NORM switch. DBK43B has two switches, CAL1/RUN and CAL2/RUN.
The switches are easy to see on the rear panel of the associated DBK module.

• The NORM and RUN positions are for normal running operations, as indicated in the following

tables.

• In the CAL position, the shunt calibration offset and the excitation voltage can be read depending on

the software function control. This is discussed in the next section.

This manual is related to the following products: