beautypg.com

Dbk90 – specifications – Measurement Computing DBK Part 2 User Manual

Page 187

background image

DBK Option Cards and Module

907492

DBK90 pg. 13

DBK90 – Specifications

Note: Specifications are subject to change without notice.

System Compatibility: Attaches to DaqBook/2000 Series, or to a WBK40 or WBK41
System Connectors: 1 male and 1 female DB37 connector for unit-to-unit mating and for mating with P1 on the primary data

acquisition device.

TC Connectors: 56 Mini-TC connectors, oriented in 4 rows of 16
ACOM (Analog Common) Connector: DBK90’s ACOM connector accepts a 2 mm banana plug (Pomona® Electronics p/n

5936-0). The ACOM connector and 2mm jack pin provide a convenient means of connecting a line to DBK90’s
analog common.

Inputs: 56 differential TC inputs, open TC detection per channel
TC Types:
J, K, T, E, S, R, B, N28, N14
Speed:
Maximum TC measurement rate is 1 ms/channel

Dimensions: 285 mm W x 88 mm D x 52 mm H (11” x 3.44” x 2.05”)
Weight: 0.96 kg (2.12 lbs)
Power Requirements: 40 mA max from ±15V; 60 mA max from +5V

DBK90 Maximum Channel Capacity

Device

Max. Channel Capacity
per Device

Max. Channel Capacity
per System

Max. DBK90 Power
Capacity per Device

DaqBook/2000
Series

896

using 16 DBK90s

3,584*

using 16 DBK90s

10 DBK90s ***

WBK40, WBK41

854

using 15 DBK90s

2,562**

using 15 DBK90s

10 DBK90s ***

* Presumes 4 DaqBook/2000 Series devices per system.
** Presumes 3 WBK40 / WBK41 devices attached to a WaveBook/516E
*** Presumes that no other active DBK modules are attached. A DBK32A power supply is necessary to power additional DBK90s

or other active DBK options.

Input Impedance: 4 MΩ (differential) in parallel with 400pF
Input Bandwidth: 1 kHz
Minimum Resolution: 0.1˚C for all TC types
TC Accuracy: Valid for one year 25˚C ambient, see following table and accuracy conditions
Operating Temperature: -20˚ to +80˚C
Relative Humidity: 0 to 95%, non-condensing
Temperature Coefficient of Accuracy: ±0.05˚C for every ˚C away from 25°C
Channel-to-Channel Crosstalk: -90 dB typ (0 to 100 Hz)
DC CMRR: -80 dB typ
AC CMRR: -80 dB typ (0 to 60 Hz)
Maximum Common Mode Voltage: ±10V
Over-Voltage Protection: ±40V

TC Accuracy

at Measurement Temperature in °C (±°C)

Type

Min

Max

-100

0

100

300

500

700

900

1100

1400

J

-200 760 0.8

0.7

0.7

0.8

0.9

0.9

— — —

K

-200 1200 0.9

0.8

0.8

0.9

1.1

1.1

1.2

1.3

T

-200 400 0.9

0.8

0.8

0.8

— — — — —

E

-270 650 0.8

0.7

0.7

0.7

0.8

— — — —

S

-50 1768 — 3.1

2.4

2.0

2.0

1.9

2.0

2.1

2.1

R

-50 1768 — 3.1

2.1

2.0

1.9

1.9

1.7

1.9

2.0

B

50

1780

— — — 4.9

3.2

2.8

2.4

2.3

2.0

N28

-270 400 1.2

0.9

0.9

0.9

— — — — —

N14

0 1300 — 0.9

0.9

0.9

1.1

1.1

1.2

1.3

Accuracy conditions:

Exclusive of thermocouple errors

Exclusive of noise

VCM = 0

This manual is related to the following products: