DaqBook/100 Series & /200 Series and DaqBoard [ISA type] Configuration
Three setup steps and needed to configure DaqBooks/100 Series & /200 Series devices and ISA-type
DaqBoards for a DBK53 [or DBK54].
1.
If not using auxiliary power, set JP1 for Analog Option Card Use, also referred to as the expanded
analog mode.
Note:
These jumpers do not
apply to /2000 Series
Devices.
To power the interface circuitry of the DBK53 or DBK54 via the internal ±15 VDC
power supply JP1 must be set to “Analog Option Card Use.” However, if using
auxiliary power, e.g., DBK32A or DBK33, you must remove both JP1 jumpers.
Refer to Power Requirements in the DBK Basics section and to the DBK32A and
DBK33 sections for more detailed information, as applicable.
2.
For DaqBook/100, DaqBook/112, and DaqBook/120 only, place the JP3 jumper in bipolar
mode.
3.
For DaqBook/100, DaqBook/112, and DaqBook/120 only, place the JP4 jumper in single-
ended mode.
Note: The DaqBook/200 Series devices and DaqBoards do not have a JP4 jumper. The single-ended or
differential choice is made via software configuration commands.
DaqBook/2000 Series and DaqBoard/2000 Series Configuration
No jumper configurations are required for the /2000 series devices.
Software Setup
Reference Notes:
o
DaqView users
-
Refer to chapter 3, DBK Setup in DaqView.
o
LogView users - Refer to chapter 4, DBK Setup in LogView.
DaqView Users: When a DBK53 or DBK54 is used with a /2000 Series Device, the Internal Clock
Speed should be set to 100 kHz as described in Chapter 3, DBK Setup in DaqView.
DBK53 and DBK54,