DBK43A & DBK43B, pg. 20

899892

DBK Option Cards and Modules

15. Calculate the LogBook or Daq device voltage/transducer units. Do this using the transducer

specifications and the total gain of the DBK43 module’s channel. Apply the units to your readings.

16. Verify the software settings by using a known load or strain and comparing the value to that

observed in DaqView’s Reading column.

Note:

Gain adjustments can be made by activating a shunt-cal resistor that is calculated to be at
maximum load.
To enable shunt-cal resistors:

(a) Select “Shunt Cal” as the Channel Type.

(b) DBK43A users: Set CAL/NORM switch to CAL.

DBK43B users: Place CAL1 and CAL2 “Up.”

(c) DBK43A users: LogView users: set the software CAL/NORM switch, in Hardware

Configuration, to “CAL.”
DBK43B users: : LogView users verify the software switch parameter is set to
1’CAL+2’CAL.

Settings can be verified via shunt-calibration.

After the final offset is made, the gain readings will be incorrect unless the circuit
offset is removed.