Sample hold [SAMPLE]

Displays and outputs the measured value at the instant of being specified by the external TIMING input.

• Retains and outputs the internal measured value at the time when the TIMING input becomes ON.

• In the following cases, the judgment standby status ("_ _ _ _" is indicated) is displayed until the first measured

value is fixed.

When you turn on the power, change the settings, change the program No., input RESET, and input Auto-

zero.

• Judgment outputs are output according to the measured value.

• The ON status of each output means that the NPN open-collector output is ON.

• The ON status of each input means that the input and the COM terminal are short-circuited.

• The RS-232C output is synchronized with the timing input when the "D-MODE" in the environment setting is set

to any option other than OFF. (

Page 8-17)

• The TIMING input and RESET input can be controlled using the RS-232C. (

Page 8-6)

• The TIMING input and the RESET input can be controlled by using the [HOLD] button and the [ESCAPE]

button respectively. (

Page 5-3)