This section describes the basic settings for
displacement measurement.
This section describes the scanning function for
stable measuring.
Scan width/interval
You can see irregularities on a zoomed in surface even
if the measurement target surface looks flat.
Depending on the irregularity, measurement cannot be
done correctly because of insufficient reflective light.
Stable measuring can be realized by making the
laser beam swing (scan) and canceling any points
that do not have enough reflective light.
■ LT-9010 (M)
■ LT-9030 (M)
SCAN
Meauring
unit
Pitch
(Interval : 1, 2, 5, 10
μm)
Measurement target
Measurement target
(Width : OFF, 4,120, 250, 500,100
μm)
Scan width
Meauring
unit
Scan width
Measurement target
(Width : OFF, 4,120, 240, 400, 560
μm)
Pitch
(Interval : 1, 2, 4, 8μm)
Basic Settings
Select a combination of scanning width and pitch
from the table shown below.
■ LT-9010 (M)
■ LT-9030 (M)
OFF
No scanning
120/2
120/5
120/10
250/2
250/5
250/10
Scan width (Width):120/250/500/1100µm
500/2
Pitch (Interval): 2/5/10 µm
500/5
Select a combination from the above.
500/10
1100/2
1100/5
1100/10
4/1
Fine-scanning: about 4 µm
OFF
No scanning
120/2
120/4
120/8
240/2
240/4
240/8
Scan width (Width):120/240/400/560µm
400/2
Pitch (Interval): 2/4/8 µm
400/4
Select a combination from the above.
400/8
560/2
560/4
560/8
4/1
Fine-scanning: about 4 µm