Changing the ranges of the areas – KEYENCE LT-9001 Series User Manual
Page 75

4
Chapter
Operations
and
Fun
ction
Settings
in
the
Profile
Mode
E LT-9001-IM
You can change the range of the vertical green
cursor (Area 1) and that of the vertical blue cursor
(Area 2) to any range.
The procedure used to change the ranges of the
areas is explained as follows.
1
Press the [ENTER] button on the light
intensity screen.
The cursor is placed on [A1].
2
Select the area (from A1 and A2) by using the
[
▲] and [▼] buttons.
[A1] is selected in this example.
3
Place the cursor on the lower limit of [A1] (the
area indicated in green) by pressing the
[ENTER] button.
The cursor moves to "–400" in this example.
4
Specify the lower limit by moving the vertical
cursor with the [t] and [s] buttons.
5
Place the cursor on the upper limit by using
the [SCREEN] button.
The cursor moves to "–300" in this example.
Changing the Ranges of the Areas
4-9
6
Specify the upper limit as well as the lower
limit.
The upper limit is changed to "–150" in this
example.
7
Press the [ENTER] button.
The area is fixed.
• You can switch the positions of the upper and lower
limits. The area that is not indicated by the cursor is
selected in such a case.
• The area setting display bar is displayed in yellow
within an area where the ranges of Area 1 and Area
2 overlap each other.
Reference