Light Intensity Screen

The value, which is set as a threshold for the

amount of light to be measured or not, is called

the light intensity level.

The light intensity level is set by using the [t] and

[s] buttons to move the vertical yellow cursor on the

light intensity screen.

On the light intensity screen, the light intensity level is

displayed in [LVL] and the received light intensity is

displayed in [INT].

Setting the Light Intensity Level

Light intensity
level

Select the measurement target surfaces from the

four highest values of light intensity.

The procedure for setting the measurement target

surface is explained as follows.

1

Select [M1] (measurement target surface 1) by

pressing the [ENTER] button on the light

intensity screen.

The cursor is placed on [M1].

2

When [MEASUREMENT] is set to [

ΔSLANT] or

[THICKNESS], select measurement target

surface 1 or 2 by using the [t] or [s] button.

When [MEASUREMENT] is set to [DISTANCE] or

[SLANT], only the measurement target surface 1

can be set and you may skip this procedure.

Refer to "MEASUREMENT" (page 3-10).

3

Select the measurement target using the [

]

and [

] buttons.

4

Press the [ENTER] button.

The measurement target surface is fixed.

Setting the Measurement Target Surface