OUTPUT Settings
The procedure used to set the LIMITS is explained as
follows.
1
Select [LIMITS] from [OUT1] or [OUT2] on the
program mode screen.
The screen appears for setting the tolerance.
2
Select [UPPER LIMIT].
3
Specify the value for [UPPER LIMIT] and
press the [ENTER] button.
The change of the upper limit is fixed.
4
Select [LOWER LIMIT].
Set the lower limit in the same manner.
The processes performed on the measured value
such as averaging, offsetting, and calibrating are
referred to as OUT settings. Each process can be
set in OUT1 and OUT2 independently.
This section describes the OUT settings common
to the displacement mode and profile mode.
Refer to "OUT Settings" (page 3-10) for the
functions unique to the displacement mode.
Refer to "OUT Settings" (page 4-12) for the
functions unique to the profile mode.
Acceptability of the measured value can be
determined. The range of measured value for this
judgment is called limits.
Set the upper and lower limits.
The relationship between the criteria of HI, GO, and
LO range and the measured value is shown below.
If the measured value is out of the tolerance range,
indications appear as follows.
"Default Values and Setting Range" (page 2-25)
Display mode
Measured value status
Measured value
display
Over the upper limit
Judgment
of tolerance
displays [HI] in
red.
Single display
The areas within the
Judgment displays
range between the
[GO] in green.
upper limit and lower limit
Over the lower limit of
Judgment displays
tolerance
[LO] in red.
Over the upper or lower
Displayed in red.
limits of the tolerance
Double display The areas within the
Displayed in
range between the
green.
upper limit and lower limit
LIMITS (Tolerance)
HI
Upper limit < Measured value
GO
Lower limit
≤ Measured value ≤ Upper limit
LO
Measured value
≤ Lower limit