KEYENCE LT-9001 Series User Manual
Page 153

9
Chapter
Specifications
E LT-9001-IM
9-5
■ When the profile output in the profile mode is set to ON
Output
Analog output
Binary output
Data output
Strobe output
Shape waveform
During RESET processing
-10.5V
-1048576
OFF
(0x100000)
During setting
Same as during measurement
Input error
Same as during measurement
Other errors
-10.5V
0 (0x000000)
OFF
When power is turned on
-10.5V
0 (0x000000)
OFF
* ON/OFF indicates the ON/OFF status of the NPN open collector output.
* RESET processing is performed after the setting change, program No. change, auto-zero processing, and
RESET input.
* Output can be performed as in the table above even during the pause screen.
During
measurement
The profile waveform
voltage is output to MON1.
The point data that exceeds
±10 V is output as +10.5 V
at the positive side and as
-10.5 V at the negative side.
The point data of dark is
-10.5 V.
The light intensity
waveform voltage is
output to MON2.
The point data that
indicates an over range
at the plus side of the
profile waveform data is
+1048575 (0x0FFFFF).
The point data that
indicates an over range
at the minus side is
-1048576 (0x100000)
The point data of dark
status is -1048576 (0 x
100000)
SCAN STB
BIN STB
* Only when I/O is
binary
pages 7-11 and 7-13
Specifications of the cable between measuring unit and controller
Model
LT-C2
LT-C10
Cable length
2 m
10 m
Weight
to 200 g
to 700 g
Note: Up to cables with a maximum total length of 20 m can be connected.
3