Auto peak hold [A.PEAK]
Measures continuously, and displays and outputs the measured value as needed.
• Auto bottom hold [A.BOTTOM]..........Displays and outputs the minimum value as needed that is measured
continuously.
• Auto peak-to-peak hold [A.P-P] .........Displays and outputs the differential between the maximum and
minimum value as needed that are measured continuously.
• Judgment outputs are output according to the measured value.
• The ON status of each output means that the NPN open-collector output is ON.
• The ON status of each input means that the input and the COM terminal are short-circuited.
• The RS-232C output is synchronized with the timing input when the "D-MODE" in the environment setting is set
to any option other than OFF. (
page 8-17)
Refer to "Binary Output" (page 7-10) for details.
• If setting the RESET input to ON when the TIMING input is ON, the judgment standby status is displayed ("- - -
-" is indicated) until the TIMING input becomes OFF.
• The TIMING input and RESET input can be controlled using the RS-232C. (
Page 8-6)
• The TIMING input and the RESET input can be controlled by using the [HOLD] button and the [ESCAPE]
button respectively. (
Page 5-3)
Reference
Judgment
Measured value
Judgment standby status