Scan – KEYENCE LT-9001 Series User Manual
Page 80
Operations
and
Fun
ction
Settings
in
the
Profile
Mode
4
Chapter
E LT-9001-IM
Basic Settings
4-14
Select a combination of scanning width an pitch from
the table shown below
The procedure used to set the width/interval is
explained as follows.
1
Select [SCAN] from [BASIC] on the program
mode screen.
The scan setting screen appears.
2
Select [WIDTH/INTERVAL].
3
Select the scanning width and the pitch by
using the [
▲] and [▼] buttons.
4
Press the [ENTER] button.
The width/interval is fixed.
120/2
120/5
120/10
250/2
250/5
Scan width (Width): 120/250/500/1100
μm
250/10
Pitch (Interval): 2/5/10
μm
500/2
Select a combination from the above.
500/5
500/10
1100/2
1100/5
1100/10
WIDTH/INTERVAL
Set the range to be measured as scanning width. Set
the pitch used to measure as [INTERVAL]. Set the
pitch in balance with the shape of the measurement
target.
• The smaller the pitch becomes, the more time that
is needed for measuring; however, the
measurement results become sharper. Considering
the stability required, set the scanning width and
pitch.
SCAN