beautypg.com

KEYENCE LT-9001 Series User Manual

Page 57

background image

3

Chapter

Operations

and

Fun

ction

Settings

in
the

Displacement

Mo

de

E LT-9001-IM

3-11

The procedure to set [MEASUREMENT] is explained

as follows.

1

Select [OUTPUT] from [OUT1] or [OUT2] on

the program mode screen.

The screen appears for setting the output.

2

Move the cursor over the items for

[MEASUREMENT] by using the [

] , [] and

[ENTER] buttons.

3

Select either [DISTANCE], [SLANT], [

ΔSLANT]

or [THCIK] by using the [

] and [] buttons.

4

Press the [ENTER] button.

[MEASUREMENT] setting is fixed.

An error may occur in the thickness measurement of

a transparent object depending on the refractive

index. In such a case, you can use the calibration

function to obtain the proper value.

Refer to "CALIB (calibration)" (page 5-16) for the

calibration function.

The [SLANT] can be set in OUT1 only and cannot be

set in OUT2.

The [SLANT] and [

ΔSLANT] can be selected only

when the scanning width is set to anything other than

[OFF] and [4/1].

Refer to "SCAN" (page 3-15) for setting scanning

width.

The measuring unit should be viewed in the direction

shown in the drawing. (+/–)

Note

Reference

Slant

The reference surface refers to the surface position

that was adjusted at the factory.

ΔSlant

You can select either surface as a reference surface.

When the target is slanted against the reference

surface in a counterclockwise direction, the value is

regarded as "+", and when the target is slanted in a

clockwise direction, the value is regarded as "–".

Thickness of transparent object (THICK)

The value is always regarded as "+".

Measuring

unit

Measurement target

Degree in relation
to the reference
surface

Measuring

unit

Measurement target

Transparent
target

Relative slant

Measuring

unit

Measuring target (transparent)

Thickness of the
transparent object