Timing details (lt-9010 (m)) – KEYENCE LT-9001 Series User Manual
Page 127

7-13
E LT-9001-IM
7
Chapter
I/O
Terminal
Sampling cycle (T2), minimum input time (internal processing time, binary strobe cycle) (T1)
Timing Details (LT-9010 (M))
[ms]
Minimum input time
Scan width/
Sampling cycle (T2)
Internal processing time (T1)
interval
OUT STB/BIN STB cycle
Displacement
Profile mode
Displacement
Profile mode
mode
FINE=OFF
FINE=ON
mode
FINE=OFF
FINE=ON
OFF
0.64
120/2
42.3
42.3
164
2.6
120/5
26.9
26.9
96
3.2
120/10
19.2
19.2
64
3.2
250/2
84.5
84.5
323
2.6
250/5
43.6
43.6
173
3.2
250/10
28.2
28.2
96
3.2
500/2
164
164
646
2.6
500/5
75.6
75.6
327
3.2
500/10
43.6
43.6
167
3.2
1100/2
356
356
1420
2.6
1100/5
153
153
711
3.2
1100/10
82
82
359
3.2
4/1
2.6
* The sampling cycle(T2) differs according to the manufacturing variations of individual measuring units.
* The sampling cycle(T2) differs according to the settings of "Scan width/interval".
* The measurement response time, average measurement time, update cycle (T5), and RESET time (T4) vary according to the
change in the sampling cycle (T2).
* The minimum input time (internal processing time, binary strobe cycle) (T1) varies according to the setting of FINE mode.
Average measurement time, RESET time (T4), update cycle (T5)
Number of
Average
Displacement mode
Profile mode
measurement times measurement Number of data for
RESET time
Update cycle Number of data for
RESET time
Update cycle
for averaging
time
moving average
(T4)
(T5)
moving average
(T4)
(T5)
2
T2x2
1
T2x3
T2x1
2
T2x5
T2x2
4
T2x4
1
T2x5
T2x1
2
T2x7
T2x2
8
T2x8
1
T2x9
T2x1
2
T2x11
T2x2
16
T2x16
1
T2x17
T2x1
2
T2x19
T2x2
32
T2x32
1
T2x33
T2x1
2
T2x35
T2x2
64
T2x64
1
T2x65
T2x1
2
T2x67
T2x2
128
T2x128
2
T2x129
T2x2
4
T2x131
T2x4
256
T2x256
4
T2x257
T2x4
512
T2x512
8
T2x513
T2x8
1024
T2x1024
16
T2x1025
T2x16
2048
T2x2048
32
T2x2049
T2x32
4096
T2x4096
64
T2x4097
T2x64
* T2 is a sampling cycle.
* Number of times for averaging raging from 256 to 4096 can only be selected when the scanning width/interval is set to OFF in the
displacement mode.
Refer to "SCAN" (page 3-15) for details.
* Average measurement time does not mean measurement response time.
Refer to "Measurement Response Time" (page 9-6)
* The moving average is used for the measured value averaging function.
Refer to "OUT (Output)" (page 5-8) for details.
0.64