HOLD MODE (Measurement mode)

Function of the hold mode

You can select how to output the measured value from the options such as outputting the measured value

along with performing the measurement or outputting the maximum or minimum value at a given interval from

the continuously changing measured value.

The LT-9001 Series has nine types of hold modes as follows.

Normal [NORMAL]

Displays and
outputs the
measured value as
needed.

Peak hold [PEAK]

Displays and
outputs the
maximum
measured value
in the specified
interval.

Bottom hold [BOTTOM]

Displays and
outputs the
minimum
measured value in
the specified
interval.

Peak-to-peak hold [P-P]

Displays and
outputs the
maximum and
minimum measured
values in the
specified interval.

Average hold [AVG.H]

Displays and
outputs the
average value of
the measured
values in the
specified interval.

Auto bottom hold [A.BOTTOM]

Displays and
outputs the
minimum value
that is measured
as needed.

Auto peak-to-peak hold [A.P-P]

Displays and
outputs the
maximum and
minimum values
that are measured
as needed.

Sample hold [SAMPLE]

Displays and
outputs the
measured value at
the specified
moment.

[ ] is what is actually
indicated on the
screen.

Auto peak hold [A.PEAK]

Displays and
outputs the
maximum
value that is
measured as
needed.

Minimum value

Minimum value

Measured value

Measured value

Measured value

Measured value

Sampling time

Sampling time

Sampling time

Sampling time

Average
value

Sample hold
value

TIMING
input

Maximum value

Maximum value

*Differential

between maximum

and minimum values

*Differential

between maximum

and minimum values

*

*