Response delay time, Display update cycle, Display update cycle response delay time – KEYENCE LT-9001 Series User Manual

Page 154: Averaging time measurement response time, Power on reset time

background image

9-6

Specifications

Averaging time

Measurement response time

The measurement response time can be calculated with the following formula.

• Distance mode:

Measurement response time = Average measurement time + Update cycle (T5) +

Internal processing time (T1)

• Profile mode:

Measurement response time = Averaging response time + Average measurement time

+ Update cycle (T5) + Internal processing time (T1)

“Timing Details” (page 7-13) for average measurement time, update cycle (T5), and internal processing

time (T1).

Power ON reset time

This is a period of time taken to start measurement after turning on the power.

• Power ON reset time = approx. 10 sec.

The update cycle of the screen display of the microscope image and measured value differs depending on the

scan settings.

■ LT-9010 (M)

■ LT-9030 (M)

Display Update Cycle

Response Delay Time

E LT-9001-IM

Specifications

9

Chapter

Number of data for averaging

Averaging time

Averaging response time

2

T2x2

T2x0

4

T2x4

T2x2

16

T2x16

T2x14

* Averaging response time is used to calculate the measurement response time.

T2 is a sampling cycle (

page 7-13).

[ms

]

Profile mode

Scan width/

Displacement

FINE=OFF

FINE=ON

Interval

mode

OFF

116

120/2

84.5

84.5

164

120/5

108

108

192

120/10

115

116

128

250/2

169

169

323

250/5

87.1

87.1

173

250/10

113

113

96

500/2

164

164

646

500/5

151

152

327

500/10

87.1

87.1

167

1100/2

356

356

1414

1100/5

153

153

711

1100/10

164

164

359

4/1

128

[ms

]

Scan width

Displacement

/Interval

mode

OFF

106

120/2

174

120/4

138

120/8

115

240/2

169

240/4

169

240/8

113

400/2

142

400/4

136

400/8

164

560/2

187

560/4

187

560/8

108

4/1

128

LT-9000-M-NO9-E.qxd 10.4.8 1:59 PM ページ9-6