Response delay time, Display update cycle, Display update cycle response delay time – KEYENCE LT-9001 Series User Manual
Page 154: Averaging time measurement response time, Power on reset time
9-6
Specifications
Averaging time
Measurement response time
The measurement response time can be calculated with the following formula.
• Distance mode:
Measurement response time = Average measurement time + Update cycle (T5) +
Internal processing time (T1)
• Profile mode:
Measurement response time = Averaging response time + Average measurement time
+ Update cycle (T5) + Internal processing time (T1)
“Timing Details” (page 7-13) for average measurement time, update cycle (T5), and internal processing
time (T1).
Power ON reset time
This is a period of time taken to start measurement after turning on the power.
• Power ON reset time = approx. 10 sec.
The update cycle of the screen display of the microscope image and measured value differs depending on the
scan settings.
■ LT-9010 (M)
■ LT-9030 (M)
Display Update Cycle
Response Delay Time
E LT-9001-IM
Specifications
9
Chapter
Number of data for averaging
Averaging time
Averaging response time
2
T2x2
T2x0
4
T2x4
T2x2
16
T2x16
T2x14
* Averaging response time is used to calculate the measurement response time.
T2 is a sampling cycle (
page 7-13).
[ms
]
Profile mode
Scan width/
Displacement
FINE=OFF
FINE=ON
Interval
mode
OFF
116
120/2
84.5
84.5
164
120/5
108
108
192
120/10
115
116
128
250/2
169
169
323
250/5
87.1
87.1
173
250/10
113
113
96
500/2
164
164
646
500/5
151
152
327
500/10
87.1
87.1
167
1100/2
356
356
1414
1100/5
153
153
711
1100/10
164
164
359
4/1
128
[ms
]
Scan width
Displacement
/Interval
mode
OFF
106
120/2
174
120/4
138
120/8
115
240/2
169
240/4
169
240/8
113
400/2
142
400/4
136
400/8
164
560/2
187
560/4
187
560/8
108
4/1
128
LT-9000-M-NO9-E.qxd 10.4.8 1:59 PM ページ9-6