Scan center
Set the scan center for scanning.
Refer to "Default Values and Setting Range"
(page 2-25) for the setting range and limitations.
The procedure used to set the scan center is
explained as follows.
1
Select [SCAN] from [BASIC] on the program
mode screen.
The scan setting screen appears.
2
Select [SCAN CENTER] and press the
[ENTER] button.
The measurement screen appears and the
center position is indicated by the vertical cursor
displayed in yellow.
3
Move the vertical cursor (yellow) to any point
by using the [t] or [s] button to set the
measurement center.
4
Press the [ENTER] button.
The scan center is fixed.
If the [ZERO] button is pressed while setting the scan
center, the parameter is set to "0".
Reference
4-15
FINE mode
Depending on the irregularity of the surface,
measurement cannot be performed correctly
because of insufficient reflective light.
FINE mode is the function used to scan the
[WIDTH/INTERVAL] (pitch) set in the scanning setting
with a greater degree of fine pitch to obtain sharper
results.
This function provides more stable measurement of
the profile waveform.
The procedure used to set the FINE mode is
explained as follows.
1
Select [SCAN] from [BASIC] on the program
mode screen.
The scan setting screen appears.
2
Set [WIDTH/INTERVAL] of the scanning.
Refer to "SCAN" (page 4-14) for setting of
[WIDTH/INTERVAL].
3
Select [FINE MODE].
4
Select [ON] by using the [
▲] and [▼] buttons.
5
Press the [ENTER] button.
FINE mode is set.
When [FINE MODE] is [ON], more accurate profile
measurement can be performed, but the button
response and measurement takes longer than usual.
Set this function considering the required degree of
accuracy.
Reference
Mark for the
maximum
scanning range
(white)
Mark for the current
scan center (white)
and scanning
width bar (orange)
Mark for the vertical
cursor (yellow)